Temporal drift in calibration of Ir-192 brachytherapy sources: A multi-center study on dosimetric discrepancies and metrological consistency

0Citations signalées, ce qui n’est pas une note de qualité
25Institutions déclarées
10Pays d’affiliation déclarés

Rattachement africain : es, dk, be, no, at, nl, gr, us, de, se. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background and purpose High-dose-rate (HDR) brachytherapy relies on accurate and metrologically traceable source strength determination. Recent anecdotal reports and preliminary studies have suggested a possible temporal drift in the reference air kerma rate ( RAKR ) measured by clinical users compared to manufacturer certificates for 192 Ir sources. This study investigates the existence and magnitude of such drift across a large, multicenter European dataset and explores potential underlying causes. Materials and methods A total of over 1700 RAKR measurements for HDR and PDR brachytherapy sources, collected over two decades from 29 centers in 10 European countries, were analyzed. The ratio of hospital-measured RAKR to manufacturer-certified RAKR was assessed using linear regression and t-tests to evaluate drift. Data were corrected for center-dependent systematics and segmented around key dates corresponding to changes in primary standards. Supplementary analyses included leave-one-out testing and time-segmented trend assessment. Results A statistically significant drift (+0.15 %/year) was detected for all 192 Ir source types after 2018, correlating temporally with updates in Physikalisch-Technische Bundesanstalt laboratory (PTB) calibration standards. Removing PTB corrections from manufacturer values nearly eliminated the observed drift. No such trend was observed for 60 Co sources. Conclusion The findings reveal a drift in hospital-to-manufacturer Ir-192 RAKR ratios that is temporally correlated with changes in metrology standards, not uniformly implemented across calibration chains. The true cause is however still unknown. The study aimed to investigate that greater transparency and harmonization among all stakeholders are essential to ensure dosimetric accuracy in HDR brachytherapy.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Temporal drift in calibration of Ir-192 brachytherapy sources: A multi-center study on dosimetric discrepancies and metrological consistency
Date Crossref
01/07/2025
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced Radiotherapy TechniquesMedical Imaging Techniques and ApplicationsRadiation Therapy and Dosimetry

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.