Accelerometer-based vibration analysis and oxygenator thrombosis in venovenous ECMO: an experimental porcine model

0Citations signalées, ce qui n’est pas une note de qualité
3Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : no. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Abstract Background Oxygenator thrombosis is a potentially life-threatening complication during venovenous extracorporeal membrane oxygenation (VV ECMO). It can cause blood flow obstruction, impaired gas exchange, hematologic abnormalities, or sudden ECMO flow cessation. Early detection and timely circuit exchange is critical yet challenging. Acute clot formation necessitates immediate circuit replacement, while premature replacement risks unnecessary procedural harm and increased costs. No reliable method exists to detect early oxygenator thrombosis. Strategies include visual inspection, monitoring the pressure difference across the oxygenator (ΔP oxy), gas exchange evaluation, and blood tests. In the present animal study, we aimed to evaluate the feasibility of accelerometer-based vibration analysis as a real-time and non-invasive method for detecting oxygenator thrombosis during VV ECMO. We hypothesized that accelerometer signals would change concurrently with or precede increases in ΔP oxy. Methods The study was performed on anesthetized and mechanically ventilated pigs (n = 7) on VV ECMO. Hemodynamic parameters, ECMO circuit pressures, and signals from an accelerometer attached to the ECMO oxygenator were continuously recorded at different pump speeds, and after anticoagulation reversal to promote thrombosis within the ECMO oxygenator. Results The primary finding of this study was a significant increase in the accelerometer signal's Root Mean Squared (RMSoxy) 15 min after anticoagulation reversal, with no rpm adjustment and without corresponding changes in ΔP oxy. Variations in RMSoxy associated with high ECMO pump speed and circuit flow were discernible from those observed following anticoagulation reversal. Conclusion The present animal study demonstrates the feasibility of accelerometer-based vibration analysis as a real-time and non-invasive method for detecting vibrations associated with reversal of anticoagulation and potential oxygenator thrombosis during VV ECMO.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Accelerometer-based vibration analysis and oxygenator thrombosis in venovenous ECMO: an experimental porcine model
Date Crossref
02/06/2025
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Mechanical Circulatory Support DevicesRespiratory Support and MechanismsCardiac Arrest and Resuscitation

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.