Standard breath-hold versus free-breathing real-time cine cardiac MRI—a prospective randomized comparison in patients with known or suspected cardiac disease
Rattachement africain : gb, au, us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Aims: Cardiovascular magnetic resonance (CMR) is established as the reference standard for cardiac volumetric assessment. Despite the accuracy and robustness of steady-state free precession (SSFP) cine imaging, its use may prove challenging in patients with arrhythmia and in those who cannot perform repeated breath holds. An alternative solution may be a free-breathing electrocardiogram (ECG)-triggered, retro-gated, real-time cine sequence. This study sought to compare left ventricular volumetric, wall motion, and thickness assessment with both techniques. Methods and results: Consecutive patients with known or suspected cardiac disease referred for clinical CMR were studied at 3-Tesla. Participants underwent short-axis standard SSFP and real-time cine imaging in a randomized order within the same scan. Between sequence agreement and mean difference were compared for end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume, ejection fraction (EF), left ventricular mass (LVM), maximal wall thickness (MWT), and wall motion score index (WMSi). Two hundred and two patients (mean age 61 ± 14 years, 51% male and 14% irregular rhythm) were studied. All left ventricular indices showed good-excellent agreement between the two methods [intraclass correlation coefficient (95% confidence interval), EDV 0.96 (0.95-0.97), ESV 0.96 (0.94-0.97), EF 0.85 (0.81-0.88), LVM 0.93 (0.91-0.95), MWT 0.80 (0.75-0.85), and WMSi 0.93 (0.91-0.95)]. Conclusion: In patients with known or suspected cardiac disease, real-time cine imaging demonstrates good-excellent reproducibility of LV volumetric, wall thickness and resting wall motion assessment when compared with standard SSFP (Trial registration: NCT05221853).
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Standard breath-hold versus free-breathing real-time cine cardiac MRI—a prospective randomized comparison in patients with known or suspected cardiac disease
- Date Crossref
- 01/01/2025
- Éditeur
- Oxford University Press (OUP)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.