Safety and Feasibility of the RevCore Catheter for Venous In-Stent Thrombosis: A Multicenter, Retrospective Study

2Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background After venous stent placement, patients may develop in-stent thrombosis (IST). The RevCore thrombectomy catheter (Inari Medical) is a novel device specifically designed to treat venous IST. Herein, the safety and feasibility of this device are evaluated. Methods Patients were retrospectively included if they were ≥18 years old, had lower extremity venous IST, and were treated using the study device. The primary end point is technical success, defined as an average postprocedural effective diameter ≥50% and calculated using stent diameter and flow channel diameter on intravascular ultrasound. The safety end points are 30-day device-related serious adverse events, namely mortality, readmission, clinically significant pulmonary embolism, and vessel perforation. Results Data from 44 patients (48 treated limbs, 6 treated stent types) from 4 centers were analyzed. Mean age was 54.8 ± 17.1 years, and 23 (52.3%) patients were female. Median IST symptom duration was 8.0 (IQR, 2.0-104.0) weeks. The primary end point was achieved in all 48 limbs. Comparing baseline with postprocedural measurements, effective stent diameter increased from 21.1% ± 26.5% to 89.6% ± 10.1%. No patient met a safety end point. On the first follow-up at 40.0 (29.0-62.5) days, primary patency was demonstrated in 32 (94.1%) of the 34 treated limbs with available data. Six (85.7%) limbs with Clinical-Etiology-Anatomy-Pathophysiology scores of C 6 at baseline had longer-term data available (3.1 ± 0.7 months), and all limbs demonstrated improvement of C 6 disease, with 3 (50.0%) improving to class C 5 . Conclusions A thrombectomy procedure for deep-vein IST reestablished patency in all cases and across multiple stent types. Patency was sustained through the first follow-up in 94% of limbs.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Safety and Feasibility of the RevCore Catheter for Venous In-Stent Thrombosis: A Multicenter, Retrospective Study
Date Crossref
01/04/2025
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Central Venous Catheters and HemodialysisVenous Thromboembolism Diagnosis and ManagementDiagnosis and Treatment of Venous Diseases

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.