Comparison of cone-beam computed tomography with photon-counting detector computed tomography for dental implant surgery
Rattachement africain : ch, nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
PURPOSE: To compare cone-beam computed tomography (CBCT) with photon-counting detector computed tomography (PCD-CT) at equivalent radiation doses, focusing on qualitative and quantitative parameters relevant to dental implant surgery. METHODS: This ex vivo comparative study of porcine specimens assessed five imaging protocols with both CBCT and PCD-CT at three effective radiation dose levels (high: 360µSv, standard: 145µSv, low: 20µSv) to evaluate image quality, artifact burden, metal artifact susceptibility, and quantitative bone measurements in the mandibular region. Three blinded readers analyzed the data using a 5-point Likert scale (5 = highest to 1 = lowest rating) and performed linear bone measurements at implant planning sites. Statistical analysis included descriptive statistics and inter-reader reliability assessment using intraclass correlation coefficients (ICC). RESULTS: Each reader evaluated 30 data sets (12 CBCT, 18 PCD-CT), with 24 implant planning sites per imaging protocol. High-dose PCD-CT demonstrated the best image quality and diagnostic interpretability (4.89 ± 0.27), followed by standard-dose PCD-CT and CBCT (4.50 ± 0.73; 4.33 ± 0.61), with low-dose protocols showing intermediate quality with higher artifact burden. In comparison to CBCT, PCD-CT demonstrated superior performance in reducing implant-induced artifacts across all protocols. Quantitative bone measurements showed minimal variability, meeting clinical precision requirements for computer-assisted implant surgery. Both qualitative (ICCs:0.70-0.89; p < 0.001) and quantitative (ICCs:0.79-1; p < 0.001) analyses demonstrated high reliability, regardless of the reader's experience. CONCLUSIONS: PCD-CT demonstrated superior image quality and reduced artifacts compared with CBCT at all radiation dose levels. These findings highlight PCD-CT's potential to enhance implant planning and improve clinical outcomes with reduced radiation exposure while maintaining diagnostic accuracy.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Comparison of cone-beam computed tomography with photon-counting detector computed tomography for dental implant surgery
- Date Crossref
- 13/03/2025
- Éditeur
- Springer Science and Business Media LLC
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.