Staying in Place: In Vitro Comparison of Extracorporeal Membrane Oxygenation Cannula Fixation for Dislodgment Prevention
Rattachement africain : at. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Background/Objectives: Secure large-bore cannula insertion is critical for effective extracorporeal membrane oxygenation (ECMO), as inadequate fixation can lead to complications such as infection, dislodgment, and life-threatening events. With inconsistent guidelines for ECMO line management, this study compares the effectiveness of traditional suture fixation to an adhesive securement method in the prevention of ECMO cannula dislodgment using an in vitro model. Methods: Porcine skin and muscle tissue sections were prepared and mounted in a custom holder. A 21F venous ECMO cannula was inserted using a modified Seldinger technique. Three fixation methods were randomly compared: (1) three silk sutures, and (2a) one silk suture with a CathGrip adhesive anchoring device. In addition, a sub-analysis was performed using (2b) the Hollister adhesive anchoring device. A uniaxial testing machine simulated 50 mm cannula dislodgment, measuring tensile forces at 12.5, 25, and 50 mm dislodgment points. Results: A total of 26 ECMO cannula fixations using sutures, 26 with adhesive CathGrip, six with a Hollister device, and three controls were tested across six porcine samples. Sutures demonstrated greater variability in force at maximum dislocation, with 27% rupturing at 50 mm. In contrast, CathGrip provided greater flexibility without tearing. The adhesive exhibited higher stiffness (2.38 N/mm vs. 2.09 N/mm, p < 0.001) and dislodgment energy (0.034 J vs. 0.032 J, p = 0.002) in the 0–5 mm range, while sutures showed greater stiffness in the 5–50 mm range (1.42 N/mm vs. 1.18 N/mm, p < 0.001). At larger displacements (25 mm and 50 mm) and in total energy absorption, no statistically significant differences were observed (p = 0.57). In a sub-analysis, the six fixations using the Hollister device exhibited higher variability and significantly lower dislodgment forces at 25 mm (p = 0.033) and 50 mm (p = 0.004) compared to the CathGrip device. Conclusions: This study suggests that adhesive anchoring methods, such as CathGrip, may provide comparable or potentially superior fixation strength to sutures for ECMO cannula stabilization under controlled conditions. However, further research, including clinical trials, is necessary to confirm these findings, evaluate long-term performance, and explore the implications for dislodgment risk and infection prevention in clinical practice.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Staying in Place: In Vitro Comparison of Extracorporeal Membrane Oxygenation Cannula Fixation for Dislodgment Prevention
- Date Crossref
- 04/03/2025
- Éditeur
- MDPI AG
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
Medical University of Vienna pays non établi dans la noticeUniversité ou école supérieure
-
Ludwig Boltzmann Institute for Cancer Research pays non établi dans la noticeStructure de recherche
-
Ludwig Boltzmann Institute for Digital Health and Prevention pays non établi dans la noticeStructure de recherche
-
Medical University Vienna Department for Cardiac and Thoracic Aortic Surgery pays non établi dans la noticeUniversité ou école supérieure
-
Ludwig Boltzmann Institute for Cardiovascular Research pays non établi dans la noticeStructure de recherche
Medical University of Vienna, Ludwig Boltzmann Institute for Cancer Research et Ludwig Boltzmann Institute for Digital Health and Prevention, avec 2 autres affiliations.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.