Aller au contenu principal
Accès ouvert déclaré 2025 article

Patient setup variation on Elekta Unity and its impact on adaptive planning

0Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

PURPOSE: The unique design of the MR-linac may restrict the use of effective immobilization devices, resulting in significant patient setup variations (PSVs). The purpose of this study is to analyze the PSVs on the Elekta Unity system and investigate their impact on adaptive planning. METHODS: The PSVs for 10 brain, 10 pancreas, five prostate, and five rectum patients previously treated on Elekta Unity were analyzed. The five prostate and five pancreas plans were selected to investigate the impact of PSVs on adaptive planning. The reference scans were shifted by 1, 2, and 3 cm in the left-right (LR) and superior-inferior (SI) directions to simulate PSVs. Both the adaptive-to-position (ATP) and adaptive-to-shape (ATS) workflows were executed. The adaptive planning time, number of monitor units (MUs), and dosimetric metrics quantifying target coverage and organ-at-risks (OARs) sparing were compared. RESULTS: For brain treatments, the average/maximum PSVs were -0.2 ± 0.3 cm/0.8 cm (LR), 0.3 ± 0.7 cm/1.8 cm (SI), and 0.8 ± 0.7 cm/1.8 cm in the anterior-posterior (AP) direction. For pancreas treatments, the PSVs are -0.1 ± 1.0 cm/3.8 cm (LR), -0.1 ± 0.8 cm/3.5 cm (SI), and 0.3 ± 0.3 cm/1.3 cm (AP). Pelvis treatments had similar PSVs as pancreas treatments. The ATS workflow took two to three times longer than the ATP workflow. The only trend observed was that the plan MUs increased slightly (< 10%) with PSVs in the ATP workflow for prostate patients. Both workflows effectively reproduced target coverage and OAR sparing, regardless of the magnitude of the PSVs. CONCLUSIONS: Significant PSVs were observed on Elekta Unity due to suboptimal patient immobilization. Using prostate and pancreas treatments as examples, we demonstrated that adaptive planning can effectively accommodate such PSVs. Nevertheless, efforts should be made to minimize PSVs-particularly rotations-to mitigate intra-fraction motion and reduce treatment time.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Patient setup variation on Elekta Unity and its impact on adaptive planning
Date Crossref
12/02/2025
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced Radiotherapy TechniquesRadiation Therapy and DosimetryCerebrospinal fluid and hydrocephalus

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.