OMERACT GCA phantom project: validation of a 3D-printed ultrasound training phantom for diagnosis of giant cell arteritis

0Citations signalées — pas une note de qualité
31Institutions déclarées
12Pays d’affiliation déclarés

Résumé fourni par la source

OBJECTIVE: Ultrasonography is crucial for diagnosing giant cell arteritis (GCA); however, training opportunities are rare. This study tested the reliability of ultrasonography findings and measurements of the intima-media thickness (IMT) among ultrasonography experts by using phantoms of the axillary (AA) and temporal arteries (TA) created with high-resolution 3D printing. METHODS: Twenty-eight participants from 12 European countries received eight sets of phantoms of the AA and the superficial TA (including common, frontal and parietal branches), which were examined in a blinded fashion according to a predefined protocol and evaluated based on Outcome Measures in Rheumatology (OMERACT) GCA ultrasound definitions. Due to difficulties with the delineation of the intima-media complex, the parietal branch of the phantoms was modified, and a second round was conducted. The IMT was measured, and phantoms were classified as normal or vasculitic. RESULTS: In both rounds, the phantoms were correctly classified as normal/abnormal in >81% of cases yielding a Fleiss' kappa of 0.80 (95% CI 0.78 to 0.81) in round 1 and 0.74 (95% CI 0.72 to 0.75) in round 2. IMT measurements revealed an intraclass correlation coefficient (ICC 1.1) of 0.98 (95% CI 0.98 to 0.99) in both rounds. Intrarater reliability was good with a median Cohens Kappa of 0.83 and median ICC of 0.78. CONCLUSION: The study demonstrated high reliability among ultrasound experts in applying the OMERACT ultrasound definitions for GCA and in measuring the IMT using a 3D-printed phantom of the AA and TA. This phantom could assist clinicians in training to assess the large arteries of patients with suspected or established GCA.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
OMERACT GCA phantom project: validation of a 3D-printed ultrasound training phantom for diagnosis of giant cell arteritis
Date Crossref
01/02/2025
Éditeur
BMJ
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Vasculitis and related conditionsSystemic Lupus Erythematosus ResearchRheumatoid Arthritis Research and Therapies

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.