Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry
Résumé fourni par la source
Abstract The Cleopatra ASIC is a 12-channel prototype ASIC for the readout of hydrogenated amorphous silicon sensors used for real-time dosimetry in radiation diagnostic and radiation therapy. The architecture is based on a current to frequency conversion based on the recycling integrator principle in order to cover a dynamic range of four orders of magnitude with high linearity. Three different input amplifier configurations have been implemented in order to check the trade-off between detector capacitance and maximum output frequency. Cleopatra has been designed in CMOS 28 nm technology and succesfully tested in laboratory.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Cleopatra: a 12-channel recycling integrator ASIC for the readout of hydrogenated amorphous silicon detectors in radiotherapy dosimetry
- Date Crossref
- 01/01/2025
- Éditeur
- IOP Publishing
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.