Multiplex lateral flow test sensitivity and specificity in detecting influenza A, B and SARS-CoV-2 in adult patients in a UK emergency department

5Citations signalées — pas une note de qualité
7Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

BACKGROUND: Rapid identification of individuals with acute respiratory infections is crucial for preventing nosocomial infections. For rapid diagnosis, especially in EDs, lateral flow devices (LFDs) are a convenient, inexpensive option with a rapid turnaround. Several 'multiplex' LFDs (M-LFDs) now exist, testing for multiple pathogens from a single swab sample. We evaluated the real-world performance of M-LFD versus PCR testing in detecting influenza A, B and SARS-CoV-2) in the ED setting. METHODS: After preliminary evaluation of an M-LFD (SureScreen) with laboratory-grown virus and PCR-negative clinical samples, it was evaluated in a real-world setting at the ED of St Thomas' Hospital (London, UK) from 1 December 2022 to 21 April 2023. Eligible participants were ≥18 years of age, admitted with respiratory symptoms and received concurrent M-LFD and PCR tests. Main endpoints were sensitivity to detect influenza A/B (primary) and SARS-CoV-2 (secondary) versus PCR. The probability of a true positive in relation to viral concentration (expressed as PCR cycle threshold (Ct)) was analysed using logistic regression. RESULTS: In total, 808 symptomatic participants were included (49.8% female; mean age 46.9 years). Test sensitivity (95% CI) was 67.0% (56.9% to 76.1%) for influenza A (n=100), 94.1% (71.3% to 99.9%) for influenza B (n=17) and 48.2% (39.7% to 56.8%) for SARS-CoV-2 (n=141). Sensitivity for SARS-CoV-2 was significantly lower than that for influenza A and B (p=0.0057 and p=0.00088, respectively). The probability of a true positive was 98% for Ct<25 for influenza A and SARS-CoV-2 (influenza B non-evaluable). No co-infections were identified by PCR or M-LFD. CONCLUSION: The real-world performance of SureScreen M-LFD was consistent with laboratory evaluation and achieved a high sensitivity for individuals with high viral concentration, most likely to be infectious. Given the representative UK population sample, results could be generalised for use in other settings.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Multiplex lateral flow test sensitivity and specificity in detecting influenza A, B and SARS-CoV-2 in adult patients in a UK emergency department
Date Crossref
14/01/2025
Éditeur
BMJ
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Respiratory viral infections researchSARS-CoV-2 detection and testingInfection Control and Ventilation

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.