5-year migration and inducible displacement of the uncemented LCS and ATTUNE rotating platform knee systems: a secondary report of a randomized controlled RSA trial
Rattachement africain : nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
BACKGROUND AND PURPOSE: Early migration of the uncemented cruciate-sacrificing rotating platform ATTUNE and Low Contact Stress (LCS) tibial components was classified as at-risk for aseptic loosening rates exceeding 6.5% at 15 years based on recent fixation-specific migration thresholds. In this secondary report of a randomized controlled trial (RCT) we aimed to evaluate whether the 5-year migration, inducible displacement, and the clinical outcome of the ATTUNE components were comparable to those of the LCS. METHODS: Patients from the initial 2-year radiostereometric analysis (RSA) RCT were recruited for a 5-year follow-up. At 5 years, participants underwent 2 supine and 1 loaded RSA examination, clinical assessments, and questionnaires. Migration was analyzed using maximum total point motion (MTPM), translations, and rotations, focusing on 5-year migration, continuous migration (> 0.10 mm/year), and inducible displacement. Revisions, along with clinical and functional outcomes, were also evaluated. RESULTS: At 5 years, 24 ATTUNE and 24 LCS implants were analyzed. The mean MTPM was similar for tibial components (ATTUNE 1.13mm [confidence interval (CI) 0.94-1.33]; LCS 1.24 mm [CI 1.05-1.46]) but significantly lower for the ATTUNE femoral component (1.14 mm [CI 0.92-1.39]) than LCS 1.87 mm [CI 1.57-2.21]). 2-to-5-year migration rates were comparable, but 11 ATTUNE and 7 LCS exceeded 0.10 mm MTPM/year, indicating a higher risk of loosening. Inducible displacement was similar, although 1 patient with a tibial ATTUNE showed excessive displacement (3.34 mm MTPM) with focal osteolysis but no symptoms. 1 revision 10 days post-surgery was performed for an ATTUNE insert spinout, resolved with an isolated insert exchange. Clinical and functional outcomes were comparable. CONCLUSION: At the 5-year follow-up, ATTUNE tibial components showed similar migration, while the femoral component migrated significantly less than the LCS, which mainly occurred during the first 2 years. 2-to-5-year migration rates, inducible displacement, and clinical and functional outcomes were comparable. These findings suggest a comparable long-term risk of aseptic loosening between the uncemented ATTUNE and LCS knee systems.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- 5-year migration and inducible displacement of the uncemented LCS and ATTUNE rotating platform knee systems: a secondary report of a randomized controlled RSA trial
- Date Crossref
- 10/01/2025
- Éditeur
- MJS Publishing, Medical Journals Sweden AB
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.