A new reference block for field gamma spectrometry applications in ESR and luminescence dating
Rattachement africain : es, au, gb, fr. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
A new granite-doped concrete block with 60 cm x 60 cm x 60 cm dimensions has been built at CENIEH, Burgos, Spain, for dosimetry calibration and cross-referencing purposes. Independent evaluations of the block’s gamma dose rate using passive Al2O3:C dosimeters and various field gamma spectrometer (NaI) probes produce consistent results of 1495 ± 51 μGy a-1 and 1514 ± 43 μGy a-1 (or 1537 ± 19 μGy a-1, depending on the evaluation procedure employed), respectively. Bulk radioelement concentrations calculated from field gamma spectrometry using the Windows method are as follows: K = 1.58 ± 0.08 %, U = 4.26 ± 0.28 ppm, and Th = 12.62 ± 0.72 ppm. This new block complements existing dosimetry reference materials accessible at other laboratories and is available for the broader trapped charge dating community to use for instrument calibration, reproducibility assessments and intercomparison studies.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- A new reference block for field gamma spectrometry applications in ESR and luminescence dating
- Date Crossref
- 11/01/2025
- Éditeur
- Cantonal and University Library Fribourg
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.