Uncomplicated Extravascular Implantable Cardioverter‐Defibrillator Extraction After 4 Years Dwell Time: A Case Report
Rattachement africain : nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
The extravascular implantable cardioverter-defibrillator (EV-ICD) was developed to overcome complications associated with transvenous leads while being able to deliver anti-tachycardia pacing (ATP). The lead is implanted in the substernal space, which makes extraction a cautious procedure. We present a case of a 51-year-old women with a successful EV-ICD extraction after a lead dwell time of 4 years, which is the longest reported. The EV-ICD lead was extracted using simple traction after the removal of all adhesions at the xiphoidal site around the lead. We advise to only use extraction tools if the initial attempt is not successful and if no adhesions at the caudal part of the lead are visible anymore, as these tools could also damage the surrounding tissue and the lead.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Uncomplicated Extravascular Implantable Cardioverter‐Defibrillator Extraction After 4 Years Dwell Time: A Case Report
- Date Crossref
- 18/12/2024
- Éditeur
- Wiley
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.