Automated Measurement of Strabismus Angle Using a Commercial Virtual Reality Headset

3Citations signalées, ce qui n’est pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : ch. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Current clinical measurements of strabismus angles (SAs) are manual and require a skilled examiner and active cooperation by patients. This results in high interexaminer variability, and clinical use is limited by the availability of trained examiners. An objective and automated procedure, independent of the examiner, would be useful. This single-center, prospective, diagnostic feasibility study compared the vertical and horizontal SA of patients, as measured with a commercially available virtual reality headset (VRH) and custom software with the gold standard measurements performed manually with the Harms tangent screen (HW) and the alternate prism cover test (ACT). We implemented the ACT by showing the patient a fixation target on each eye alternatively with the VRH while recording the eye position of the patient. We then processed the data with custom written software to calculate the SA of the patients. These measurements were then compared to the SA measured with HW and the clinical ACT. Thirty-three patients took part in our study. We found good correlation between the VRH method and the HW as well as the clinical ACT. Best correlation was found for horizontal SA in the primary position, with the vertical SA in the primary position also correlating well. Peripheral gaze resulted in a slightly lower correlation due to the overestimation of horizontal SA and underestimation of vertical SA; cyclorotation was not measured with the VRH. Overall, VRH, HW, and clinical ACT correlated similarly well as the published interexaminer correlation for ACT. The automated measurement of strabismus with a VRH is feasible, easily applicable, fast, accurate, and can be run on consumer hardware that is affordable and increasingly available.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Automated Measurement of Strabismus Angle Using a Commercial Virtual Reality Headset
Date Crossref
18/12/2024
Éditeur
Georg Thieme Verlag KG
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Ophthalmology and Eye DisordersGlaucoma and retinal disordersVestibular and auditory disorders

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.