Cross-Field Strength and Multi-Vendor Validation of MagDensity for MRI-based Quantitative Breast Density Analysis
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Abstract Purpose Breast density (BD) is a significant risk factor for breast cancer, yet current assessment methods lack automation, quantification, and cross-platform consistency. This study aims to evaluate MagDensity, a novel magnetic resonance imaging (MRI)-based quantitative BD measure, for its validity and reliability across different imaging platforms. Methods Ten healthy volunteers participated in this prospective study, undergoing fat-water MRI scans on three scanners: 3T Siemens Prisma, 3T Siemens Biograph mMR, and 1.5T GE Signa. Great effort was made to schedule all scans within a narrow three-hour window on the same day to minimize any potential intraday variations, highlighting the logistical challenges involved. BD was assessed using the MagDensity technique, which included combining magnitude and phase images, applying a fat-water separation technique, employing an automated whole-breast segmentation algorithm, and quantifying the volumetric water fraction. The agreement between measures was analyzed using mean differences, two-tailed t-tests, Pearson’s correlation coefficients, and Bland-Altman plots. Results No statistically significant differences in BD measurements by MagDensity within the same field strength and vendor (3T Siemens), with high correlation (Pearson’s r > 0.99) and negligible mean differences (< 0.2%). Cross-platform comparison between the 3T Siemens and the 1.5T GE scanners showed mean differences of < 5%. After linear calibration, these variations were reduced to insignificant levels, yielding a strong correlation (Pearson’s r > 0.97) and mean differences within ±0.2%. Conclusion MagDensity, an MRI-based BD measure, exhibits robustness and reliability across diverse scanner models, vendors, and field strengths, marking a promising advancement towards standardizing BD measurements across multiple MRI platforms. It provides a valuable tool for monitoring subtle longitudinal changes in BD, which is vital for breast cancer prevention and personalized treatment strategies.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Cross-Field Strength and Multi-Vendor Validation of MagDensity for MRI-based Quantitative Breast Density Analysis
- Date Crossref
- 10/12/2024
- Éditeur
- openRxiv
- Type
- posted-content
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
Emory University pays non établi dans la noticeUniversité ou école supérieure
-
Stony Brook University pays non établi dans la noticeUniversité ou école supérieure
Emory University et Stony Brook University.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.