A low-cost setup for orientation-dependent post-mortem MRI under temperature control
Le résumé fourni par la source
Motivation: Systematic variation of external parameters can provide insight into whether theoretical models appropriately describe MRI contrast. Goal(s): Our goal was to develop a cost-effective setup for comprehensive temperature- and orientation-dependent relaxation and magnetization-transfer experiments in post-mortem tissue on a clinical scanner. Approach: A remotely tiltable Helmholtz coil was integrated into a thermally insulated box, where the temperature can be adjusted by a heated airflow. Results: Robust coil performance, accurate adjustment of the sample orientation relative to B0 (±1°), and stable temperature conditions (±0.5 °C) were achieved. Theoretically expected temperature dependencies of T1 and diffusivity in agarose were experimentally reproduced. Impact: Well-defined variations of the temperature (between ambient temperature and 45 °C) and sample orientation (between 0 and 90° relative to B0) in MRI experiments with small post-mortem tissue specimens were achieved on a clinical scanner with a cost-effective setup.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- A low-cost setup for orientation-dependent post-mortem MRI under temperature control
- Date Crossref
- 26/11/2024
- Éditeur
- ISMRM
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.