A fast, vendor-neutral protocol for multi-center, multi-parametric quantitative MRI studies in brain tumor patients
Le résumé fourni par la source
Motivation: Multi-centric multi-parametric quantitative MRI (mp-qMRI) studies require short and vendor-neutral protocols. Goal(s): The goal was to develop and validate an 8-minute, vendor-neutral protocol for multi-center mp-qMRI studies on brain tumor patients at 3T. Approach: 5 Volunteers were measured at two scanners. Application of the proposed method was demonstrated for one brain metastasis patient, where artefact free qMRI maps were obtained. Results: qMRI maps (T1, T2*, PD and QSM) obtained from 5 volunteers on two scanners showed a very good reproducibility. Among the parameters, PD yielded the lowest COV. Artefact free qMRI maps are demonstrated in a brain metastasis patient. Impact: We propose an 8-minute, vendor-neutral mp-qMRI protocol for 3T studies in brain tumor patients. T1, T2*, PD and QSM maps are demonstrated in 5 healthy volunteers and one brain metastsis patient. Multi-centric qMRI studies could use this fast mp-qMRI protocol.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- A fast, vendor-neutral protocol for multi-center, multi-parametric quantitative MRI studies in brain tumor patients
- Date Crossref
- 26/11/2024
- Éditeur
- ISMRM
- Type
- proceedings-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.