Abstract Su502: Is 80% a Reasonable Minimal Goal for Defibrillation Efficacy of Out-of-Hospital Cardiac Arrest ?

0Citations signalées — pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

Introduction: There is no clear goal defined by standards or learned societies on clinical performance related to defibrillation efficacy. In literature, monophasic shocks defibrillation efficacy is mainly >70% whereas biphasic is >80%. Resuscitation guidelines define efficient defibrillation as cessation of shockable rhythm at 5s post-shock and proceed to state that beyond 5s, shock outcomes are biased by known confounders (call to AED connection time (CallAEDTime), bystander, cardio-pulmonary resuscitation (CPR), drugs…). Thus, it is consistent that automated external defibrillator (AED) performance should be assessed in terms of defibrillation efficacy right after the shock rather than in terms of restoration of a spontaneous circulation or survival to hospital discharge. This study aimed to define an acceptable goal for validation of defibrillation efficacy. Method: In this observational retrospective study, we included AED tracings of adult non-traumatic out-of-hospital cardiac arrest (OHCA) patients with initial ventricular fibrillation (VF) rescued by basic life support teams (Paris Fire Brigade) from 01/2017 to 02/2022. Cardiac rhythm was examined during the 10 first minutes of connection of the AED (DGT7, Schiller Médical, France). Shockable rhythm is defined as coarse-VF with at least 200µV peak-to-peak amplitude (ptpA). Efficient defibrillation was defined as cessation of shockable rhythm at 5s post-shock. We fitted a linear mixed model to predict defibrillation efficacy using the 4s ptpA of pre-shock VF. The model included the patient as random effect and a p-value <0.05 was considered statistically significant. Results: We analyzed 1333 shocks from 547 patients, with: median age of 59 (interquartile range, IQR[49-71]), 78% male, 36% in public places, 72% with bystander CPR and median[IQR] CallAEDTime of 13min [10-15]min. The effect of the pre-shock ptpA of VF on the defibrillation efficacy is statistically significant: OR=1.023, 95% Confidence Interval (CI) [1.010,1.036], p<0.001. That means that for each 100µv increment in VF amplitude pre-shock, the probability of success at 5s post-shock increases by 2.3%. Adjusting for age, sex, location, CPR and CallAEDTime does not change the results. The logistic regression of Figure1, shows a defibrillation efficacy at 200µV ptpA of 79%, CI [75,82]%. Conclusion: Based on 1333 OHCA biphasic shocks, this study suggests defining 80% as a reasonable minimal goal for validation of the AED defibrillation efficacy.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Abstract Su502: Is 80% a Reasonable Minimal Goal for Defibrillation Efficacy of Out-of-Hospital Cardiac Arrest ?
Date Crossref
12/11/2024
Éditeur
Ovid Technologies (Wolters Kluwer Health)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Cardiac Arrest and ResuscitationRespiratory Support and MechanismsMechanical Circulatory Support Devices

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.