Clinical precision of Aoralscan 3 and Emerald S on the palatal and dentition areas: Evaluation for forensic applications
Rattachement africain : hu. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
OBJECTIVES: To compare the precision (repeatability and intermediate precision) of palatal and dentition scans taken with two different intraoral scanners. METHODS: The maxillary arch of 23 individuals was scanned three times using the Aoralscan 3 (Shining 3D) and Emerald S (Planmeca), resulting in 6 scans per individual. The scans were segmented in Meshmixer into palatal and dentition areas. Each replicate of an individual was compared within the specific scanner (repeatability) and to the corresponding replicate of the other scanner (intermediate precision). The scans were aligned using the iterative closest-point algorithm in the Zeiss Inspect software. The mean absolute distance between the aligned surfaces was calculated. Statistical comparisons were made using Friedman's two-way analysis of variance. Data are presented in median (quartile 1; quartile 3) form. RESULTS: No significant difference in repeatability was found between Emerald S and Aoralscan 3 for the palate (26 µm [22; 26] vs. 22 µm [18; 26]) and for dentition (37 µm [31; 44] vs. 38 µm [35; 48]. Intermediate precision of the palate (32 µm [26; 43]) and dentition (72 µm [63; 80]) was significantly lower than the repeatability of Emerald S (p < 0.05, p < 0.001) and of Aorlascan 3 (p < 0.001, p < 0.001). Both precision types of both scanners were significantly lower for the dentition than for the palate (p < 0.001). CONCLUSION: The precision of the dentition scan is lower than that of the palate. The precision of the Aoralscan 3 is similar to that of the Emerald S scanner. CLINICAL SIGNIFICANCE: Novel intraoral scanners could be used with high precision for palatal soft tissue scans, expanding their clinical utility. Aligning scans from two different IOSs still has high precision, facilitating the interchangeable use of intraoral scanners for orthodontic, prosthetic, and forensic examinations.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Clinical precision of Aoralscan 3 and Emerald S on the palatal and dentition areas: Evaluation for forensic applications
- Date Crossref
- 01/02/2025
- Éditeur
- Elsevier BV
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.