Inhalation through two mini-spacer-like inhaler technique training devices: An in-vitro aerosol delivery comparative study

0Citations signalées — pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

Introduction: Patients are recommended to inhale slowly (30 L/min) over 5 sec through pressurized metered dose inhalers (pMDIs)1. Poor pMDI use is common among patients. Flo-Tone® (FT) and Flo-Tone controlled resistance® (FTCR) (Clement Clarke International, UK) are mini-spacer pMDI technique training devices with audible feedback to guide patients to a co-ordinated inhalation through pMDIs. Methods: 30 and 90 L/min inhalation flow rates over 2 and 5 sec were used to compare aerosol delivery of salbutamol from Ventolin Evohaler connected to FT and FTCR. The Next Generation Impactor (NGI) was used to evaluate the delivered dose (DD), fine particle fraction (FPF), mass median aerodynamic diameter (MMAD). Deposition on FT, FTCR and NGI’s induction port (IP) were also compared. Results: Comparing FT to FTCR, no significant differences in DD except at 90 L/min over 5 sec (88.2 vs 71.2 µg). The MMAD was significantly different at only 30 L/min over 5 sec (2.44 vs 3.0 µm) and 90 L/min over 2 sec (2.42 vs 2.17). The FPF was similar. FTCR retained less drug and resulted in less IP impaction. Conclusion: The FT and FTCR delivered comparable salbutamol doses and fine particles (≤5 µm) under optimal and suboptimal inhalation through pMDI. The FTCR was privileged with lower throat deposition. Ultimately, these devices are intended to improve and maintain patients’ correct pMDI use. References: 1. Laube, B., Janssens, H., de Jongh, F et al. (2011). Eur Respir J 37, 1308–1331. doi: 10.1183/09031936.00166410

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Inhalation through two mini-spacer-like inhaler technique training devices: An in-vitro aerosol delivery comparative study
Date Crossref
14/09/2024
Éditeur
European Respiratory Society
Type
proceedings-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Inhalation and Respiratory Drug Delivery

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.