Device–device interference triggered by an abandoned pacemaker: a case report
Rattachement africain : de. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Background: Cardiac implantable electronic devices (CIEDs) are prone to electromagnetic interference. Common sources include household electronics and industrial machinery. However, medical equipment can also trigger interferences and cause CIED malfunction. Case summary: We report on a 79-year-old male with sudden onset of presyncope. He has a long history of device therapy, including an active VVI leadless pacemaker and an abandoned abdominal pacemaker with epicardial leads. Automatic reactivation of the abandoned pacemaker due to reaching end-of-life mode led to interaction with the active pacemaker, inhibiting it in its function. Due to elevated capture threshold and insufficient output, pacing by the reactivated pacemaker was accompanied by intermittent loss of myocardial capture and patient symptoms. By changing the mode to VOO and increasing the pacing rate of the active pacemaker, the interaction was prevented as an intermittent solution. The final therapy consisted of explanting the pulse generator of the abdominal pacemaker. Discussion: We present a patient with a deactivated abandoned cardiac pacemaker, which self-activated after reaching end-of-life mode and triggered an interaction with his active pacemaker. This case emphasizes the importance of explanting old devices to avoid potential interaction.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Device–device interference triggered by an abandoned pacemaker: a case report
- Date Crossref
- 01/11/2024
- Éditeur
- Oxford University Press (OUP)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
Medizinische Hochschule Hannover pays non établi dans la noticeUniversité ou école supérieure
-
Hannover Heart Rhythm Center pays non établi dans la noticeInstitution
Medizinische Hochschule Hannover et Hannover Heart Rhythm Center.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.