Ventilator performances for non-invasive ventilation: a bench study

7Citations signalées — pas une note de qualité
9Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

Introduction A wide range of recent ventilators, dedicated or not, is available for non-invasive ventilation (NIV) in respiratory or intensive care units (ICU). We conducted a bench study to compare their technical performances. Methods Ventilators, including five ICU ventilators with NIV mode on, two dedicated NIV ventilators and one transport ventilator, were evaluated on a test bench for NIV, consisting of a 3D manikin head connected to an ASL 5000 lung model via a non-vented mask. Ventilators were tested according to three simulated lung profiles (normal, obstructive, restrictive), three levels of simulated air leakage (0, 15, 30 L/min), two levels of pressure support (8, 14 cmH2O) and two respiratory rates (15, 25 cycles/min). Results The global median Asynchrony Index (AI) was higher with ICU ventilators than with dedicated NIV ventilators (4% (0; 76) vs 0% (0; 15), respectively; p<0.05) and different between all ventilators (p<0.001). The AI was higher with ICU ventilators for the normal and restrictive profiles (p<0.01) and not different between ventilators for the obstructive profile. Auto-triggering represented 43% of all patient-ventilator asynchrony. Triggering delay, cycling delay, inspiratory pressure-time product, pressure rise time and pressure at mask were different between all ventilators (p<0.01). Dedicated NIV ventilators induced a lower pressure-time product than ICU and transport ventilators (p<0.01). There was no difference between ventilators for minute ventilation and peak flow. Conclusion Despite the integration of NIV algorithms, most recent ICU ventilators appear to be less efficient than dedicated NIV ventilators. Technical performances could change, however, according to the underlying respiratory disease and the level of air leakage.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Ventilator performances for non-invasive ventilation: a bench study
Date Crossref
01/10/2024
Éditeur
BMJ
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Respiratory Support and MechanismsNosocomial Infections in ICUMechanical Circulatory Support Devices

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.