Diagnostic Accuracy of Beta‐2 Transferrin Gel Electrophoresis for Detecting Cerebrospinal Fluid Rhinorrhea

8Citations signalées, ce qui n’est pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

OBJECTIVE: Unilateral thin clear rhinorrhea (UTCR) may represent a variety of pathologies including cerebrospinal fluid (CSF) rhinorrhea. Beta-2 transferrin (B2Tf) gel electrophoresis (GE) has become the preferred testing modality due to reportedly high sensitivity (87%-100%) and specificity (71%-100%). However, there have been relatively few studies assessing its diagnostic accuracy. The purpose of this single-institution study was to determine the accuracy of B2Tf GE in detecting CSF rhinorrhea. METHODS: A single-center retrospective review was conducted from 2016 and 2024 for all patients who presented with UTCR and underwent B2Tf GE. Institutional review board approval was obtained. The gold standard for diagnostic confirmation of true and false positives (TP, FP) as well as false negatives (FN) was endoscopic exploration. The gold standard for true negative (TN) was response to medical therapy. RESULTS: A total of 105 patients underwent 149 B2Tf GE tests. 40 (38.1%) patients were diagnosed with CSF rhinorrhea. Of the 149 B2-Tf GE tests, there were 51 TPs, 72 TNs, 20 FPs, and 6 FNs yielding 89.5% sensitivity, 78.3% specificity, 71.8% positive predictive value, and 92.3% negative predictive value, respectively. Of the false results the most common causes for error were purulent sinusitis (n = 6, 23.1%), possible mucous contamination from nose-blowing during collection (n = 3, 11.5%), patient collection error (n = 3, 11.5%), and blood contamination (n = 1, 3.8%). CONCLUSION: Although these single-institutional data demonstrate test accuracy within ranges previously reported in the literature, they also demonstrate diagnostic limitations. Future studies should explore reasons for erroneous B2Tf GE results and how these may change clinical decision-making. LEVEL OF EVIDENCE: 4 Laryngoscope, 135:94-96, 2025.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Diagnostic Accuracy of Beta‐2 Transferrin Gel Electrophoresis for Detecting Cerebrospinal Fluid Rhinorrhea
Date Crossref
14/10/2024
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Head and Neck Surgical OncologySinusitis and nasal conditionsCerebrospinal fluid and hydrocephalus

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.