Analytical evaluation of eight qualitative FIT for haemoglobin products, for professional use in the UK

0Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : gb. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

OBJECTIVES: Qualitative faecal immunochemical tests for haemoglobin (FIT), for triaging for colorectal cancer investigations, are available for professional use. The aim was to evaluate these lateral flow tests. No previous analytical evaluations have been published. METHODS: Analytical sensitivity (AS) was assessed using samples spanning manufacturers' quoted AS, concurrently with the quantitative OC-SENSOR PLEDIA, using Hb-spiked samples in manufacturers' buffer (n≥5; ≤9-99 ng Hb/mL buffer), Hb-spiked feces (n=6; <2-34 µg Hb/g feces) and natural feces (n=17; <2-82 μg/g); concentrations for 50 %/100 % Hb-detected were compared with quoted AS. Compatibility with two external quality assessment schemes (EQAS) (n=9; 3-96 μg/g) and prozone compared with manufacturers limits (n=9; 2,500-10,000,000 ng/mL) were tested. Ease-of-use by five healthcare personnel was assessed. RESULTS: Seven products showed lower AS (ng/mL) than manufacturers quoted using Hb-spiked aqueous samples compared with OC-SENSOR, one was equivocal; six manufacturers quoted AS in µg/g, five showed lower AS using Hb-spiked feces. Results were similar but less consistent for natural feces. Result lines for low concentrations can be faint and open to interpretation. Results were consistent with manufacturers quoted prozone limits. Results were consistent for seven products for two EQAS. The ease-of-use was 68.5-85.6 %; products with lower scores could be improved with better instructions and sample bottles. CONCLUSIONS: AS was lower for seven products (aqueous samples) and five products (aqueous/faecal samples) and prozone consistent with manufacturers expected concentrations, compared with OC-SENSOR. EQAS results were mostly consistent with expected results; products can be used by healthcare professionals, though some manufacturer improvements could be made.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Analytical evaluation of eight qualitative FIT for haemoglobin products, for professional use in the UK
Date Crossref
09/10/2024
Éditeur
Walter de Gruyter GmbH
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Colorectal Cancer Screening and DetectionClinical Laboratory Practices and Quality ControlBiosensors and Analytical Detection

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.