Introduction and Validation of Low-Cost Ocular Biometry in Healthy and Cataractous Eyes Using a Thermally Tunable Swept-Source Laser

1Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : at, de. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

PURPOSE: To introduce and validate a novel substantially lower-priced and rapid swept-source investigational optical biometer in healthy and cataractous eyes, using a thermally tuned laser diode used extensively in cell phones and data communication as an alternative swept source. DESIGN: Prospective accuracy, validity, and reliability analysis. METHODS: A total of 60 eyes of 59 subjects (29 eyes of 29 healthy subjects and 31 eyes of 30 cataract patients) were enrolled in a prospective comparative study at the Vienna General Hospital between August 2021 and April 2023. Averaged intraocular distances were acquired in 2.5 seconds from datasets consisting of 5000 consecutive A-scans at a single position by a low-cost swept-source optical biometry (SSOB) system. Instrument repeatability was assessed via standard deviations (SDs) and coefficients of variation (CoVs) of parameters such as axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and central corneal thickness (CCT). Healthy subjects and cataract patients were subsequently measured on the same day with the SSOB and a referential partial coherence interferometry (PCI) biometer (IOLMaster 500) to establish AL inter-device correlation (r) for instrument calibration. AL and ACD as shared parameters between both biometers were evaluated for their limits of agreements (LoA) using Bland-Altman analysis. RESULTS: Repeated measurements of AL, ACD, LT, and CCT revealed SDs of 18 µm, 12 µm, 12 µm, and 10 µm, respectively. All parameters except for CCT had a COV <1%. Except for 1 eye with white cataract, 59 eyes of 59 study participants with various degrees and types of cataract could be measured with both devices. The AL inter-device correlation was excellent (r > 0.99). The 95% LoAs between both biometers were -0.14 to 0.13 mm for AL and -0.28 to 0.25 mm for ACD. CONCLUSIONS: Optical biometry using a thermally tunable VCSEL swept-source light source has the potential to provide clinically relevant biometric parameters at an unprecedented 100-fold lower price point than currently used state-of-the-art optical biometers, paving the way for compact devices in remote care settings.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Introduction and Validation of Low-Cost Ocular Biometry in Healthy and Cataractous Eyes Using a Thermally Tunable Swept-Source Laser
Date Crossref
01/01/2025
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Ophthalmology and Visual Impairment StudiesGlaucoma and retinal disordersOptical Coherence Tomography Applications

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.