Photon-Counting Detector CT Radiological-Histological Correlation in Cadaveric Human Lung Nodules and Airways

12Citations signalées, ce qui n’est pas une note de qualité
4Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : jp. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

OBJECTIVES: The aim of this study was to compare the performances of photon-counting detector computed tomography (PCD-CT) and energy-integrating detector computed tomography (EID-CT) for visualizing nodules and airways in human cadaveric lungs. MATERIALS AND METHODS: Previously obtained 20 cadaveric lungs were scanned, and images were prospectively acquired by EID-CT and PCD-CT at a radiation dose with a noise level equivalent to the diagnostic reference level. PCD-CT was scanned with ultra-high-resolution mode. The EID-CT images were reconstructed with a 512 matrix, 0.6-mm thickness, and a 350-mm field of view (FOV). The PCD-CT images were reconstructed at 3 settings: PCD-512: same as EID-CT; PCD-1024-FOV350: 1024 matrix, 0.2-mm thickness, 350-mm FOV; and PCD-1024-FOV50: 1024 matrix, 0.2-mm thickness, 50-mm FOV. Two specimens per lung were examined after hematoxylin and eosin staining. The CT images were evaluated for nodules on a 5-point scale and for airways on a 4-point scale to compare the histology. The Wilcoxon signed rank test with Bonferroni correction was performed for statistical analyses. RESULTS: Sixty-seven nodules (1321 μm; interquartile range [IQR], 758-3105 μm) and 92 airways (851 μm; IQR, 514-1337 μm) were evaluated. For nodules and airways, scores decreased in order of PCD-1024-FOV50, PCD-1024-FOV350, PCD-512, and EID-CT. Significant differences were observed between series other than PCD-1024-FOV350 versus PCD-1024-FOV50 for nodules (PCD-1024-FOV350 vs PCD-1024-FOV50, P = 0.063; others P < 0.001) and between series other than EID-CT versus PCD-512 for airways (EID-CT vs PCD-512, P = 0.549; others P < 0.005). On PCD-1024-FOV50, the median size of barely detectable nodules was 604 μm (IQR, 469-756 μm) and that of barely detectable airways was 601 μm (IQR, 489-929 μm). On EID-CT, that of barely detectable nodules was 837 μm (IQR, 678-914 μm) and that of barely detectable airways was 1210 μm (IQR, 674-1435 μm). CONCLUSIONS: PCD-CT visualized small nodules and airways better than EID-CT and improved with high spatial resolution and potentially can detect submillimeter nodules and airways.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Photon-Counting Detector CT Radiological-Histological Correlation in Cadaveric Human Lung Nodules and Airways
Date Crossref
20/08/2024
Éditeur
Ovid Technologies (Wolters Kluwer Health)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced X-ray and CT ImagingAdvanced X-ray Imaging TechniquesRadiation Dose and Imaging

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.