Echo planar imaging (EPI) is well-suited for temperature mapping in the mid-field (0.5T) thanks to longer T2* values that can be leveraged without prohibitively slow update rates. In this work, we optimize EPI based PRF thermometry acquisitions for mid-field MRI. Digital simulations and in-vivo temperature stabilities were computed over echo times ranging from 50ms – 100ms. The optimal mean temperature uncertainty over the entire acquisition volume was measured to be approximately 1⁰C demonstrating that EPI based thermometry provides clinically viable sensitivity at 0.5T with sufficient image resolution and extremely fast update rates (1.3-2 s).

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Optimization of EPI based PRF thermometry at 0.5 T
Date Crossref
14/08/2024
Éditeur
ISMRM
Type
proceedings-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les sujets associés

Atomic and Subatomic Physics ResearchAdvanced MRI Techniques and ApplicationsPhotoacoustic and Ultrasonic Imaging

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.