Time-of-flight PET/CT suppresses CT based attenuation correction and scatter coincidence correction errors due to misalignment of the gastrointestinal tract.

0Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : jp. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Objectives: This study aimed to examine the influence of changes in CT values on PET images, specifically focusing on errors in CT-based attenuation correction and scatter coincidence correction (CTAC/SC) caused by gastrointestinal gas. Furthermore, it aimed to demonstrate the effectiveness of time-of-flight (TOF) PET in reducing CTAC/SC errors. Methods: PET images were reconstructed using multiple CT images with varying CT values. The study then compared the fluctuations in pixel values of the PET images corresponding to the different CT values utilized for CTAC/SC between non-TOF and TOF acquisitions. Results: PET pixel values fluctuated with changes in CT values. In the phantom study, TOF showed a significantly smaller change in PET pixel value of 1.00±0.27 kBq/mL compared to 3.72±1.33 kBq/mL in the non-TOF at sites with a CT change of +1000 HU. In the patient study, a linear regression analysis was performed to determine the effect of changes in CT values due to gastrointestinal gas migration on standard uptake value (SUV).The results showed that the TOF group had a lower ratio of change in SUV to change in CT values compared to the non-TOF group. These findings revealed that PET pixel values exhibited fluctuations in response to changes in CT values, and TOF-PET effectively mitigated CTAC/SC errors arising from gastrointestinal gas. Conclusions: TOF-PET has the potential to reduce the occurrence of suspicious accumulation.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

La source scientifique ouverte est momentanément indisponible.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Medical Imaging Techniques and ApplicationsRadiation Dose and ImagingAdvanced X-ray and CT Imaging

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.