Impact of patient habitus and acquisition protocol on iodine quantification in dual-source photon-counting computed tomography

4Citations signalées — pas une note de qualité
2Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

PurposeEvaluation of iodine quantification accuracy with varying iterative reconstruction level, patient habitus, and acquisition mode on a first-generation dual-source photon-counting computed tomography (PCCT) system.ApproachA multi-energy CT phantom with and without its extension ring equipped with various iodine inserts (0.2 to 15.0 mg/ml) was scanned over a range of radiation dose levels (CTDIvol 0.5 to 15.0 mGy) using two tube voltages (120, 140 kVp) and two different source modes (single-, dual-source). To assess the agreement between nominal and measured iodine concentrations, iodine density maps at different iterative reconstruction levels were utilized to calculate root mean square error (RMSE) and generate Bland–Altman plots by grouping radiation dose levels (ultra-low: <1.5; low: 1.5 to 5; medium: 5 to 15 mGy) and iodine concentrations (low: <5; high: 5 to 15 mg/mL).ResultsOverall, quantification of iodine concentrations was accurate and reliable even at ultra-low radiation dose levels. RMSE ranged from 0.25 to 0.37, 0.20 to 0.38, and 0.25 to 0.37 mg/ml for ultra-low, low, and medium radiation dose levels, respectively. Similarly, RMSE was stable at 0.31, 0.28, 0.33, and 0.30 mg/ml for tube voltage and source mode combinations. Ultimately, the accuracy of iodine quantification was higher for the phantom without an extension ring (RMSE 0.21 mg/mL) and did not vary across different levels of iterative reconstruction.ConclusionsThe first-generation PCCT allows for accurate iodine quantification over a wide range of iodine concentrations and radiation dose levels. Stable accuracy across iterative reconstruction levels may allow further radiation exposure reductions without affecting quantitative results.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Impact of patient habitus and acquisition protocol on iodine quantification in dual-source photon-counting computed tomography
Date Crossref
26/07/2024
Éditeur
SPIE-Intl Soc Optical Eng
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Advanced X-ray and CT ImagingMedical Imaging Techniques and ApplicationsRadiation Dose and Imaging

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.