Metrology of two wearable sleep trackers against polysomnography in patients with sleep complaints

13Citations signalées — pas une note de qualité
11Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

Sleep trackers are used widely by patients with sleep complaints, however their metrological validation is often poor and relies on healthy subjects. We assessed the metrological validity of two commercially available sleep trackers (Withings Activité/Fitbit Alta HR) through a prospective observational monocentric study, in adult patients referred for polysomnography (PSG). We compared the total sleep time (TST), REM time, REM latency, nonREM1 + 2 time, nonREM3 time, and wake after sleep onset (WASO). We report absolute and relative errors, Bland-Altman representations, and a contingency table of times spent in sleep stages with respect to PSG. Sixty-five patients were included (final sample size 58 for Withings and 52 for Fitbit). Both devices gave a relatively accurate sleep start time with a median absolute error of 5 (IQR -43; 27) min for Withings and -2.0 (-12.5; 4.2) min for Fitbit but both overestimated TST. Withings tended to underestimate WASO with a median absolute error of -25.0 (-61.5; -8.5) min, while Fitbit tended to overestimate it (median absolute error 10 (-18; 43) min. Withings underestimated light sleep and overestimated deep sleep, while Fitbit overestimated light and REM sleep and underestimated deep sleep. The overall kappas for concordance of each epoch between PSG and devices were low: 0.12 (95%CI 0.117-0.121) for Withings and VPSG indications 0.07 (95%CI 0.067-0.071) for Fitbit, as well as kappas for each VPSG indication 0.07 (95%CI 0.067-0.071). Thus, commercially available sleep trackers are not reliable for sleep architecture in patients with sleep complaints/pathologies and should not replace actigraphy and/or PSG.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Metrology of two wearable sleep trackers against polysomnography in patients with sleep complaints
Date Crossref
14/06/2024
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Obstructive Sleep Apnea ResearchSleep and related disordersSleep and Wakefulness Research

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.