Aller au contenu principal
Accès ouvert déclaré 2024 article

Can you feel the force just right? Tactile force feedback for training of minimally invasive surgery—evaluation of vibration feedback for adequate force application

5Citations signalées, ce qui n’est pas une note de qualité
7Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : de, no. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: Tissue handling is a crucial skill for surgeons and is challenging to learn. The aim of this study was to develop laparoscopic instruments with different integrated tactile vibration feedback by varying different tactile modalities and assess its effect on tissue handling skills. METHODS: Standard laparoscopic instruments were equipped with a vibration effector, which was controlled by a microcomputer attached to a force sensor platform. One of three different vibration feedbacks (F1: double vibration > 2 N; F2: increasing vibration relative to force; F3: one vibration > 1.5 N and double vibration > 2 N) was applied to the instruments. In this multicenter crossover trial, surgical novices and expert surgeons performed two laparoscopic tasks (Peg transfer, laparoscopic suture, and knot) each with all the three vibration feedback modalities and once without any feedback, in a randomized order. The primary endpoint was force exertion. RESULTS: A total of 57 subjects (15 surgeons, 42 surgical novices) were included in the trial. In the Peg transfer task, there were no differences between the tactile feedback modalities in terms of force application. However, in subgroup analysis, the use of F2 resulted in a significantly lower mean-force application (p-value = 0.02) among the student group. In the laparoscopic suture and knot task, all participants exerted significantly lower mean and peak forces using F2 (p-value < 0.01). These findings remained significant after subgroup analysis for both, the student and surgeon groups individually. The condition without tactile feedback led to the highest mean and peak force exertion compared to the three other feedback modalities. CONCLUSION: Continuous tactile vibration feedback decreases the mean and peak force applied during laparoscopic training tasks. This effect is more pronounced in demanding tasks such as laparoscopic suturing and knot tying and might be more beneficial for students. Laparoscopic tasks without feedback lead to increased force application.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Can you feel the force just right? Tactile force feedback for training of minimally invasive surgery—evaluation of vibration feedback for adequate force application
Date Crossref
04/06/2024
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Surgical Simulation and TrainingTactile and Sensory InteractionsSoft Robotics and Applications

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.