Quasi-static mechanical evaluation of canine cementless total hip replacement broaches: effect of tooth design on broach and stem insertion

2Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Abstract Background Biomedtrix BFX ® cementless total hip replacement (THR) requires the use of femoral broaches to prepare a press-fit envelope within the femur for subsequent stem insertion. Current broaches contain teeth that crush and remove cancellous bone; however, they are not particularly well-suited for broaching sclerotic (corticalized) cancellous bone. In this study, three tooth designs [Control, TG1 (additional V-grooves), TG2 (diamond tooth pattern)] were evaluated with a quasi-static testing protocol and polyurethane test blocks simulating normal and sclerotic bone. To mimic clinical broaching, a series of five sequential broach insertions were used to determine cumulative broaching energy (J) and peak loads during broach insertion. To determine the effect of broach tooth design on THR stem insertion, a BFX ® stem was inserted into prepared test blocks and insertion and subsidence energy and peak loads were determined. Results Broach tooth design led to significant differences in broaching energy and peak broaching loads in test blocks of both densities. In low density test blocks, TG1 required the lowest cumulative broaching energy (10.76 ±0.29 J), followed by Control (12.18 ±1.20 J) and TG2 (16.66 ±0.78 J) broaches. In high density test blocks, TG1 required the lowest cumulative broaching energy (32.60 ±2.54 J) as compared to Control (33.25 ±2.16 J) and TG2 (59.97 ±3.07 J). During stem insertion and subsidence testing, stem insertion energy for high density test blocks prepared with Control broaches was 14.53 ± 0.81 J, which was significantly lower than blocks prepared with TG1 (22.53 ± 1.04 J) or TG2 (19.38 ± 3.00 J) broaches. For stem subsidence testing in high density blocks, TG1 prepared blocks required the highest amount of energy to undergo subsidence (14.49 ± 0.49 J), which was significantly greater than test blocks prepared with Control (11.09 ±0.09 J) or TG2 (12.57 ± 0.81 J) broaches. Conclusions The additional V-grooves in TG1 broaches demonstrated improved broaching performance while also generating press-fit envelopes that were more resistant to stem insertion and subsidence. TG1 broaches may prove useful in the clinical setting; however additional studies that more closely simulate clinical broach impaction are necessary prior to making widespread changes to THR broaches.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Quasi-static mechanical evaluation of canine cementless total hip replacement broaches: effect of tooth design on broach and stem insertion
Date Crossref
24/05/2024
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Orthopaedic implants and arthroplastyDental materials and restorationsEndodontics and Root Canal Treatments

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.