Radiation Exposure and Contrast Agent Use during Endovascular Aortic Repair Using Mobile Versus Fixed Angiography Systems

2Citations signalées, ce qui n’est pas une note de qualité
3Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : de. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background: For (thoracic) endovascular aortic repair ((T)EVAR) procedures, both mobile (standard operating room (SOR)) and fixed C-arm (hybrid operating room (HOR)) systems are available. This study evaluated differences in key procedural parameters, and procedural success for (T)EVAR in the SOR versus the HOR. Methods: All patients who underwent standard elective (T)EVAR at the Clinic for Vascular and Endovascular Surgery at the University Hospital Duesseldorf, Germany, between 1 January 2012 and 1 January 2019 were included. Data were retrieved from archived medical records. Endpoints were analyzed for SOR versus HOR during (T)EVAR. Results: A total of 93 patients, including 50 EVAR (SOR (n = 20); HOR (n = 30)) and 43 TEVAR (SOR (n = 22); HOR (n= 21)) were included. The dose area product (DAP) for EVAR and TEVAR was lower in the SOR than in the HOR (EVAR, SOR: 1635 ± 1088 cGy·cm2; EVAR, HOR: 7819 ± 8928 cGy·cm2; TEVAR, SOR: 8963 ± 34,458 cGy·cm2; TEVAR, HOR: 14,591 ± 11,584 cGy·cm2 (p < 0.05)). Procedural fluoroscopy time was shorter in the SOR than in the HOR for EVAR and TEVAR (EVAR, SOR: 7 ± 4 min; EVAR, HOR: 18.8 ± 11.3 min; TEVAR, SOR: 6.6 ± 9.6 min; TEVAR, HOR: 13.9 ± 11.8 min (p < 0.05)). Higher volumes of contrast agent were applied during EVAR and TEVAR in the SOR than in the HOR (EVAR, SOR: 57.5 ± 20 mL; EVAR: HOR: 33.3 ± 5 mL (p < 0.05); TEVAR; SOR: 71.5 ± 53.4 mL, TEVAR, HOR: 48.2 ± 27.5 mL (p ≥ 0.05). Conclusion: The use of a fixed C-arm angiography system in the HOR results in higher radiation exposure and longer fluoroscopy times but lower contrast agent volumes when compared with mobile C-arm systems in the SOR. Because stochastic radiation sequelae are more likely to be tolerated in an older patient population and, in addition, there is a higher incidence of CKD in this patient population, allocation of patients to the HOR for standard (T)EVAR seems particularly advisable based on our results.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Radiation Exposure and Contrast Agent Use during Endovascular Aortic Repair Using Mobile Versus Fixed Angiography Systems
Date Crossref
29/02/2024
Éditeur
MDPI AG
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Aortic aneurysm repair treatmentsRadiation Dose and ImagingAortic Disease and Treatment Approaches

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.