Noninvasive monitoring of the vagus nerve during thyroid surgery using cutaneous adhesive and needle electrodes: What is the optimal configuration?
Rattachement africain : us, cn. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
OBJECTIVE: Endotracheal tube (ETT) surface electrodes are used to monitor the vagus nerve (VN), recurrent laryngeal nerve (RLN), and external branch of the superior laryngeal nerve (EBSLN) during thyroid and parathyroid surgery. Alternative nerve monitoring methods are desirable when intubation under general anesthesia is not desirable or possible. In this pilot study, we compared the performance of standard ETT electrodes to four different noninvasive cutaneous recording electrode types (two adhesive electrodes and two needle electrodes) in three different orientations. METHODS: The VN was stimulated directly during thyroid and parathyroid surgery using a Prass stimulator probe. Electromyographic (EMG) responses for each patient were recorded using an ETT plus one of the following four cutaneous electrode types: large-foot adhesive, small-foot adhesive, long-needle and short-needle. Each of the four electrode types was placed in three orientations: (1) bilateral, (2) ipsilateral mediolateral, and (3) ipsilateral craniocaudal. RESULTS: Four surgical cases were utilized for data collection with the repetitive measures obtained in each subject. Bilateral electrode orientation was superior to ipsilateral craniocaudal and ipsilateral mediolateral orientations. Regardless of electrodes type, all amplitudes in the bilateral orientation were >100 μV. When placed bilaterally, the small-foot adhesive and the long-needle electrodes obtained the highest EMG amplitudes as a percentage of ETT amplitudes. CONCLUSION: Cutaneous electrodes could potentially be used to monitor the VN during thyroid and parathyroid procedures. Different electrode types vary in their ability to record amplitudes and latencies. Bilateral orientation improves EMG responses in all electrode types. Additional validation of cutaneous electrodes as an alternative noninvasive method to monitor the VN is needed.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Noninvasive monitoring of the vagus nerve during thyroid surgery using cutaneous adhesive and needle electrodes: What is the optimal configuration?
- Date Crossref
- 31/01/2024
- Éditeur
- Wiley
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.