Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with International Organization for Standardization (ISO) 17511:2020 standard: communication from the International Federation of Clinical Chemistry and Laboratory Medicine–SSC/ISTH working group on prothrombin time/international normalized ratio standardization

23Citations signalées, ce qui n’est pas une note de qualité
17Institutions déclarées
7Pays d’affiliation déclarés

Rattachement africain : nl, no, gb, it, de, us, jp. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Calibration of prothrombin time (PT) in terms of international normalized ratio (INR) has been outlined in "Guidelines for thromboplastins and plasmas used to control oral anticoagulant therapy" (World Health Organization, 2013). The international standard ISO 17511:2020 presents requirements for manufacturers of in vitro diagnostic (IVD) medical devices (MDs) for documenting the calibration hierarchy for a measured quantity in human samples using a specified IVD MD. The objective of this article is to define an unequivocal, metrologically traceable calibration hierarchy for the INR measured in plasma as well as in whole blood samples. Calibration of PT and INR for IVD MDs according to World Health Organization guidelines is similar to that in cases where there is a reference measurement procedure that defines the measurand for value assignment as described in ISO 17511:2020. We conclude that, for PT/INR standardization, the optimal calibration hierarchy includes a primary process to prepare an international reference reagent and measurement procedure that defines the measurand by a value assignment protocol conforming to clause 5.3 of ISO 17511:2020. A panel of freshly prepared human plasma samples from healthy adult individuals and patients on vitamin K antagonists is used as a commutable secondary calibrator as described in ISO 17511:2020. A sustainable metrologically traceable calibration hierarchy for INR should be based on an international protocol for value assignment with a single primary reference thromboplastin and the harmonized manual tilt tube technique for clotting time determination. The primary international reference thromboplastin reagent should be used only for calibration of successive batches of the secondary reference thromboplastin reagent.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with International Organization for Standardization (ISO) 17511:2020 standard: communication from the International Federation of Clinical Chemistry and Laboratory Medicine–SSC/ISTH working group on prothrombin time/international normalized ratio standardization
Date Crossref
01/04/2024
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Clinical Laboratory Practices and Quality ControlAnalytical Methods in PharmaceuticalsHemodynamic Monitoring and Therapy

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.