Vascular Complications Following Transcatheter Aortic Valve Implantation, Using MANTA (Collagen Plug-Based) versus PROSTAR (Suture-Based), from a French Single-Center Retrospective Registry

5Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : fr. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

TAVI requires a large-bore arteriotomy. Closure is usually performed by the suture system. Some studies report a vascular complication rate of up to 21%. MANTA is a recently developed percutaneous closure system dedicated to large caliber vessels based on an anchoring system. Early studies report a lower rate of vascular complications with MANTA devices. This single-center retrospective study included all patients who underwent femoral TAVI at the Brest University Hospital from 20 November 2019 to 31 March 2021. The primary endpoint is the rate of vascular complications (major and minor) pre and post-TAVI procedure. In total, 264 patients were included. There were no significant differences in vascular complications (major and minor) between the two groups (13.6% in the MANTA group versus 21.2% in the PROSTAR group; p = 0.105), although there was a tendency to have fewer minor vascular complications in the Manta group (12.1% versus 20.5%; p = 0.067). Manta was associated with a lower rate of bleeding complications (3.8% versus 15.2%; p = 0.002), predominantly minor complications with fewer closure failures (4.5% versus 13.6%; p = 0.01), less use of covered stents (4.5% versus 12.9%; p = 0.016), and with no difference in the need for vascular surgery compared to the Prostar group (1.5% versus 2.3%; p = 0.652). On the other hand, Manta was associated with a higher rate of femoral stenosis (4.5% versus 0%; p = 0.013) without clinical significance (1.5% versus 0%; p = 0.156). The Manta and Prostar devices are equivalent in terms of vascular complications. The Manta, compared to the Prostar, is associated with fewer bleeding complications.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Vascular Complications Following Transcatheter Aortic Valve Implantation, Using MANTA (Collagen Plug-Based) versus PROSTAR (Suture-Based), from a French Single-Center Retrospective Registry
Date Crossref
23/10/2023
Éditeur
MDPI AG
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Cardiac Valve Diseases and TreatmentsInfective Endocarditis Diagnosis and ManagementVascular Procedures and Complications

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.