NHS-POx-loaded patch versus fibrin sealant patch in a porcine robotic liver bleeding model

2Citations signalées — pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

BACKGROUND: The management of bleeding is paramount to any surgical procedure. With the increased use of less invasive laparoscopic and robotic methods, achieving hemostasis can be challenging since the surgeons cannot manually apply hemostatic agents directly onto bleeding tissue. In this study, we assessed the use of a pliable hemostatic sealant patch comprising fibrous gelatin carrier impregnated with poly(2-oxazoline) (NHS-POx) for hemostasis in robotic liver resection in a porcine bleeding model. METHODS: The NHS-POx-loaded patch (GATT-Patch), was first evaluated in a Feasibility Study to treat surgical bleeding in 10 lesions, followed by a Comparative Study in which the NHS-POx patch was compared to a standard-of-care fibrin sealant patch (TachoSil), in 36 lesions (superficial, resection, or deep injuries mimicking metastasectomies). For each lesion type, the NHS-POx and fibrin sealant patches were used in an alternating fashion with 18 lesions treated with NHS-POx and 18 with the fibrin patch. Animal preparation and surgical procedures were consistent across studies. The primary outcome was time to hemostasis (TTH) within 3 min for the Feasibility Study and within 5 min for the Comparative Study. RESULTS: In the Feasibility Study, 8 of the 10 NHS-POx-treated lesions achieved hemostasis at 30 s and 3 min. In the Comparative Study, all 18 NHS-POx patch-treated lesions and 9 of the 18 fibrin sealant patch-treated lesions achieved hemostasis at 5 min. Median TTH with NHS-POx vs fibrin sealant patch was 30 vs 300 s (P < 0.001). CONCLUSIONS: In this animal study, hemostasis during robotic liver surgery was achieved faster and more often with the NHS-POx loaded vs fibrin sealant patch.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
NHS-POx-loaded patch versus fibrin sealant patch in a porcine robotic liver bleeding model
Date Crossref
28/08/2023
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Hemostasis and retained surgical itemsHepatocellular Carcinoma Treatment and PrognosisMinimally Invasive Surgical Techniques

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.