Aller au contenu principal
Accès ouvert déclaré 2023 article

Decreasing inconsistent alarms notifications: a pragmatic clinical trial in a post-anesthesia care unit

1Citations signalées — pas une note de qualité
7Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

BACKGROUND: Alarms alert healthcare professionals of deviations from normal/physiologic status. However, alarm fatigue may occur when their high pitch and diversity overwhelm clinicians, possibly leading to alarms being disabled, paused, and/or ignored. We aimed to determine whether a staff educational program on customizing alarm settings of bedside monitors may decrease inconsistent alarms in the Post-Anesthesia Care Unit (PACU). METHODS: This is a prospective, analytic, quantitative, pragmatic, open-label, single-arm study. The outcome was evaluated on PACU admission before (P1) and after (P2) the implementation of the educational program. The heart rate, blood pressure, and oxygen saturation alarms were selected for clinical consistency. RESULTS: A total of 260 patients were included and 344 clinical alarms collected, with 270 (78.4%) before (P1), and 74 (21.6%) after (P2) the intervention. Among the 270 alarms in P1, 45.2% were inconsistent (i.e., false alarms), compared to 9.4% of the 74 in P2. Patients with consistent alarms occurred in 30% in the P1 and 27% in the P2 (p = 0.08). Patients with inconsistent alarms occurred in 25.4% in the P1 and in 3.8% in the P2. Ignored consistent alarms were reduced from 21.5% to 2.6% (p = 0.004) in the P2 group. The educational program was a protective factor for the inconsistent clinical alarm (OR = 0.11 [95% CI 0.04-0.3]; p < 0.001) after adjustments for age, gender, and ASA physical status. CONCLUSION: Customizing alarm settings on PACU admission proved to be a protective factor against inconsistent alarm notifications of multiparametric monitors.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Decreasing inconsistent alarms notifications: a pragmatic clinical trial in a post-anesthesia care unit
Date Crossref
01/05/2024
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Healthcare Technology and Patient MonitoringIntensive Care Unit Cognitive DisordersNon-Invasive Vital Sign Monitoring

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.