Aller au contenu principal
Accès ouvert déclaré 2023 preprint

Diagnostic performance of LumiraDx SARS-CoV-2 rapid antigen test compared to PCR for diagnosis of COVID-19 in Liberia

0Citations signalées — pas une note de qualité
6Institutions déclarées
2Pays d’affiliation déclarés

Résumé fourni par la source

Abstract Background Diagnosis of SARS-CoV-2 infection in Liberia has primarily relied on polymerase chain reaction (PCR)-based testing at the country’s National Reference Laboratory. This centralized approach caused reporting delays, prompting evaluation of point-of-care antigen-based tests. We assessed the test performance of the LumiraDx™ SARS-CoV-2 Ag test (LumiraDx™, London, UK) in this setting. Methods We tested ambulatory individuals screened for enrollment into an observational cohort study of COVID-19 sequelae in Monrovia in 2021. We compared the results of LumiraDx testing of anterior nasal swab specimens to the results of PCR BioFire® R2.1P (bioMérieux, Salt Lake City, Utah) on eluent from nasopharyngeal swabs. Results We evaluated 348 individuals. Among the 274 persons with symptoms, 49.3% were PCR-positive and 36.5% were antigen-positive. The sensitivity, specificity, negative predictive value (NPV), and positive predictive value (PPV) of LumiraDx in this group were 72.6% (95% CI: 64.3%-79.9%), 98.6% (95% CI: 94.9%-99.8%), 78.7% (CI: 71.9%-84.6%), and 98.0% (CI: 93.0%-99.8%), respectively. Among the 74 asymptomatic individuals, 12.2% were positive by PCR, and 5.5% by antigen testing, resulting in a sensitivity, specificity, NPV, and PPV of LumiraDx of 44.4% (95% CI: 13.7%-78.8%), 100% (95% CI: 94.5%-100%), 92.9% (CI: 84.1%-97.6%), and 100% (CI: 39.8%-100%), respectively. Conclusion Although the specificity and PPV of LumiraDx for diagnosing SARS-CoV-2 were high among persons with and without symptoms, the sensitivity was unacceptably low in both groups, and much less than that reported by the manufacturer. Before new point-of-care diagnostics are adopted, test performance needs to be assessed in the local setting.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Diagnostic performance of LumiraDx SARS-CoV-2 rapid antigen test compared to PCR for diagnosis of COVID-19 in Liberia
Date Crossref
06/08/2023
Éditeur
openRxiv
Type
posted-content

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

SARS-CoV-2 detection and testingSARS-CoV-2 and COVID-19 ResearchCOVID-19 diagnosis using AI

BNTIC News n’est pas le producteur de ces données. Exploration à la demande auprès d’OpenAlex, avec contrôle bibliographique public par Crossref. Aucun service payant requis, aucune réponse conservée. Sources et limites.