Validation of a portable monitor compared with polysomnography for screening of obstructive sleep apnea in polio survivors

2Citations signalées, ce qui n’est pas une note de qualité
8Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : cn, us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Subjective Sleep-disordered breathing (SDB) is highly prevalent in polio survivors. Obstructive sleep apnea (OSA) is the most frequent type. Full polysomnography (PSG) is recommended for OSA diagnosis in patients with comorbidities by current practice guidelines, but it is not always accessible. The purpose of this study was to evaluate whether type 3 portable monitor (PM) or type 4 PM might be a viable alternative to PSG for the diagnosis of OSA in postpolio subjects. Methods A total of 48 community-living polio survivors (39 men and 9 women) with an average age of 54.4 ± 5.3 years referred for the evaluation of OSA and who volunteered to participate were recruited. First, they completed the Epworth Sleepiness Scale (ESS) questionnaire and underwent pulmonary function testing and blood gas tests the day before PSG night. Then, they underwent an overnight in-laboratory PSG with a type 3 PM and type 4 PM recording simultaneously. Results The AHI from PSG, respiratory event index (REI) from type 3 PM, and ODI3 from type 4 PM was 30.27 ± 22.51/h vs. 25.18 ± 19.11/h vs. 18.28 ± 15.13/h, respectively (P < 0.001). For AHI ≥ 5/h, the sensitivity and specificity of REI were 95.45 and 50%, respectively. For AHI ≥ 15/h, the sensitivity and specificity of REI were 87.88% and 93.33%, respectively. The Bland–Altman analysis of REI on PM vs. AHI on PSG showed a mean difference of −5.09 (95% confidence interval [CI]: −7.10, −3.08; P < 0.001) with limits of agreement ranging from −18.67 to 8.49 events/h. ROC curve analysis for patients with REI ≥ 15/h showed an area under the curve (AUC) of 0.97. For AHI ≥ 5/h, the sensitivity and specificity of ODI3 from type 4 PM were 86.36 and 75%, respectively. For patients with AHI ≥ 15/h, the sensitivity was 66.67%, and the specificity was 100%. Conclusion Type 3 PM and Type 4 PM could be alternative ways to screen OSA for polio survivors, especially for moderate to severe OSA.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Validation of a portable monitor compared with polysomnography for screening of obstructive sleep apnea in polio survivors
Date Crossref
09/05/2023
Éditeur
Frontiers Media SA
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Sleep and related disordersViral Infections and Immunology ResearchObstructive Sleep Apnea Research

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.