Military Standard Testing of Commercially Available Supraglottic Airway Devices for Use in a Military Combat Setting

5Citations signalées, ce qui n’est pas une note de qualité
0Institutions déclarées
0Pays d’affiliation déclarés

Le résumé fourni par la source

INTRODUCTION: Airway obstruction is the second leading cause of death on the battlefield. The harsh conditions of the military combat setting require that devices be able to withstand extreme circumstances. Military standards (MIL-STD) testing is necessary before devices are fielded. We sought to determine the ability of supraglottic airway (SGA) devices to withstand MIL-STD testing. METHODS: We tested 10 SGA models according to nine MIL-STD-810H test methods. We selected these tests by polling five military and civilian emergency-medicine subject matter experts (SMEs), who weighed the relevance of each test. We performed tests on three devices for each model, with operational and visual examinations, to assign a score (1 to 10) for each device after each test. We calculated the final score of each SGA model by averaging the score of each device and multiplying that by the weight for each test, for a possible final score of 2.6 to 26.3. RESULTS: The scores for the SGA models were LMA Classic Airway, 25.9; AuraGain Disposable Laryngeal Mask, 25.5; i-gel Supraglottic Airway, 25.2; Solus Laryngeal Mask Airway, 24.4; LMA Fastrach Airway, 24.4; AuraStraight Disposable Laryngeal Mask, 24.1; King LTS-D Disposable Laryngeal Tube, 22.1; LMA Supreme Airway, 21.0; air-Q Disposable Intubating Laryngeal Airway, 20.1; and Baska Mask Supraglottic Airway, 18.1. The limited (one to three) samples available for testing provide adequate preliminary information but restrict the range of failures that could be discovered. CONCLUSIONS: Lower scoring SGA models may not be optimal for military field use. Models scoring sufficiently close to the top performers (LMA Classic, AuraGain, i-gel, Solus, LMA Fastrach, AuraStraight) may be viable for use in the military setting. The findings of our testing should help guide device procurement appropriate for different battlefield conditions.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Military Standard Testing of Commercially Available Supraglottic Airway Devices for Use in a Military Combat Setting
Date Crossref
01/01/2023
Éditeur
Breakaway Media, LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les sujets associés

Airway Management and Intubation TechniquesVoice and Speech DisordersCardiac Arrest and Resuscitation

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.