Commissioning and performance evaluation of commercially available mobile imager for image guided total body irradiation
Rattachement africain : jp, us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
BACKGROUND: The setup of lung shield (LS) in total body irradiation (TBI) with the computed radiography (CR) system is a time-consuming task and has not been quantitatively evaluated. The TBI mobile imager (TBI-MI) can solve this problem through real-time monitoring. Therefore, this study aimed to perform commissioning and performance evaluation of TBI-MI to promote its use in clinical practice. METHODS: The source-axis distance in TBI treatment, TBI-MI (CNERGY TBI, Cablon Medical B.V.), and the LS position were set to 400, 450, and 358 cm, respectively. The evaluation items were as follows: accuracy of image scaling and measured displacement error of LS, image quality (linearity, signal-to-noise ratio, and modulation transfer function) using an EPID QC phantom, optimal thresholding to detect intra-fractional motion in the alert function, and the scatter radiation dose from TBI-MI. RESULTS: The accuracy of image scaling and the difference in measured displacement of the LS was <4 mm in any displacements and directions. The image quality of TBI imager was slightly inferior to the CR image but was visually acceptable in clinical practice. The signal-to-noise ratio was improved at high dose rate. The optimal thresholding value to detect a 10-mm body displacement was determined to be approximately 5.0%. The maximum fraction of scattering radiation to irradiated dose was 1.7% at patient surface. CONCLUSION: MI-TBI can quantitatively evaluate LS displacement with acceptable image quality. Furthermore, real-time monitoring with alert function to detect intrafraction patient displacement can contribute to safe TBI treatment.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Commissioning and performance evaluation of commercially available mobile imager for image guided total body irradiation
- Date Crossref
- 26/12/2022
- Éditeur
- Wiley
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.