Magnetic resonance imaging reveals possible cause of diplopia after Baerveldt glaucoma implantation
Rattachement africain : nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
PURPOSE: To assess if ocular motility impairment, and the ensuing diplopia, after Baerveldt Glaucoma device (BGI) implantation, is related to the presence of a large fluid reservoir (bleb), using Magnetic Resonance Imaging (MRI). METHODS: In a masked observational study (CCMO-registry number: NL65633.058.18), the eyes of 30 glaucoma patients with (n = 12) or without diplopia (n = 18) who had previously undergone BGI implantation were scanned with a 7 Tesla MRI-scanner. The substructures of the BGI-complex, including both blebs and plate, were segmented in 3D. Primary outcomes were a comparison of volume and height of the BGI-complex between patients with and without diplopia. Comparisons were performed by using an unpaired t-test, Fisher's Exact or Mann-Whitney test. Correlations were determined by using Spearman correlation. RESULTS: The median volume and height of the BGI-complex was significantly higher in patients with compared to patients without diplopia (p = 0.007 and p = 0.025, respectively). Six patients had an excessively large total bleb volume (median of 1736.5mm3, interquartile range 1486.3-1933.9mm3), four of whom experienced diplopia (33% of the diplopia patients). Fibrotic strands through the BGI plate, intended to limit the height of the bleb, could be visualized but were not related to diplopia (75% versus 88%; p = 0.28). CONCLUSIONS: With MRI, we show that in a significant number of diplopia cases a large bleb is present in the orbit. Given the large volume of these blebs, they are a likely explanation of the development of diplopia in at least some of the patients with diplopia after BGI implantation. Additionally, the MR-images confirm the presence of fibrotic strands. As these strands are also visible in patients with a large bleb, they are apparently not sufficient to restrict the bleb height.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Magnetic resonance imaging reveals possible cause of diplopia after Baerveldt glaucoma implantation
- Date Crossref
- 20/10/2022
- Éditeur
- Public Library of Science (PLoS)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
Rotterdam Eye Hospital Department of Glaucoma pays non établi dans la noticeÉtablissement de santé
-
Leiden University Medical Center Department of Ophthalmology pays non établi dans la noticeOrganisme public
-
Rotterdam Ophthalmic Institute pays non établi dans la noticeStructure de recherche
Department of Glaucoma — Rotterdam Eye Hospital, Department of Ophthalmology — Leiden University Medical Center et Rotterdam Ophthalmic Institute.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.