Nuclear Recoil Calibration at Sub-keV Energies in LUX and Its Impact on Dark Matter Search Sensitivity
LUX Collaboration, D. S. Akerib, S. Alsum, H. M. Araújo, X. Bai, J. Balajthy, J. Bang, A. Baxter, E. P. Bernard, A. Bernstein, T. P. Biesiadzinski, E. M. Boulton, B. Boxer, P. Brás, S. Burdin, D. Byram, M. C. Carmona-Benitez, C. Chan, J. E. Cutter, L. de Viveiros, E. Druszkiewicz, A. Fan, S. Fiorucci, R. J. Gaitskell, C. Ghag, M. G. D. Gilchriese, C. B. Gwilliam, C. Hall, S. J. Haselschwardt, S. A. Hertel, D. P. Hogan, M. Horn, D. Q. Huang, C. M. Ignarra, R. G. Jacobsen, O. Jahangir, W. Ji, K. Kamdin, K. Kazkaz, D. Khaitan, E. V. Korolkova, S. Kravitz, V. A. Kudryavtsev, E. Leason, K. T. Lesko, J. Liao, J. Lin, A. Lindote, I. Lopes, A. Manalaysay, R. L. Mannino, N. Marangou, D. N. McKinsey, D. -M. Mei, J. A. Morad, A. St. J. Murphy, A. Naylor, C. Nehrkorn, H. Nelson, F. Neves, A. Nilima, K. C. Oliver-Mallory, K. J. Palladino, C. Rhyne, Q. Riffard, G. R. C. Rischbieter, P. Rossiter, S. Shaw, T. A. Shutt, C. Silva, M. Solmaz, V. N. Solovov, P. Sørensen, T. J. Sumner, N. Swanson, M. Szydagis, D. J. Taylor, R. Taylor, WC Taylor, B. P. Tennyson, P. A. Terman, D. R. Tiedt, W. To, L. Tvrznikova, U. Utku, A. Vacheret, Audrius Vaitkus, V. Velan, R. C. Webb, James T. White, T. J. Whitis, M. S. Witherell, F. L. H. Wolfs, D. Woodward, X. Xiang, Jing Xu, C. Zhang
5Citations signalées, ce qui n’est pas une note de qualité
29Institutions déclarées
3Pays d’affiliation déclarés
Rattachement africain : us, gb, pt.
Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Dual-phase xenon time projection chamber (TPC) detectors offer heightened sensitivities for dark matter detection across a spectrum of particle masses. To broaden their capability to low-mass dark matter interactions, we investigated the light and charge responses of liquid xenon (LXe) to sub-keV nuclear recoils. Using neutron events from a pulsed Adelphi Deuterium-Deuterium neutron generator, an in situ calibration was conducted on the LUX detector. We demonstrate direct measurements of light and charge yields down to 0.45 keV and 0.27 keV, respectively, both approaching single quanta production, the physical limit of LXe detectors. These results hold significant implications for the future of dual-phase xenon TPCs in detecting low-mass dark matter via nuclear recoils.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
La source scientifique ouverte est momentanément indisponible.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.
Les sujets associés
Dark Matter and Cosmic PhenomenaParticle Detector Development and PerformanceParticle physics theoretical and experimental studies