Open-Source Automated Insulin Delivery in Type 1 Diabetes

180Citations signalées, ce qui n’est pas une note de qualité
6Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : nz, cz. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: Open-source automated insulin delivery (AID) systems are used by many patients with type 1 diabetes. Data are needed on the efficacy and safety of an open-source AID system. METHODS: In this multicenter, open-label, randomized, controlled trial, we assigned patients with type 1 diabetes in a 1:1 ratio to use an open-source AID system or a sensor-augmented insulin pump (control). The patients included both children (defined as 7 to 15 years of age) and adults (defined as 16 to 70 years of age). The AID system was a modified version of AndroidAPS 2.8 (with a standard OpenAPS 0.7.0 algorithm) paired with a preproduction DANA-i insulin pump and Dexcom G6 CGM, which has an Android smartphone application as the user interface. The primary outcome was the percentage of time in the target glucose range of 70 to 180 mg per deciliter (3.9 to 10.0 mmol per liter) between days 155 and 168 (the final 2 weeks of the trial). RESULTS: A total of 97 patients (48 children and 49 adults) underwent randomization (44 to open-source AID and 53 to the control group). At 24 weeks, the mean (±SD) time in the target range increased from 61.2±12.3% to 71.2±12.1% in the AID group and decreased from 57.7±14.3% to 54.5±16.0% in the control group (adjusted difference, 14 percentage points; 95% confidence interval, 9.2 to 18.8; P<0.001), with no treatment effect according to age (P = 0.56). Patients in the AID group spent 3 hours 21 minutes more in the target range per day than those in the control group. No severe hypoglycemia or diabetic ketoacidosis occurred in either group. Two patients in the AID group withdrew from the trial owing to connectivity issues. CONCLUSIONS: In children and adults with type 1 diabetes, the use of an open-source AID system resulted in a significantly higher percentage of time in the target glucose range than the use of a sensor-augmented insulin pump at 24 weeks. (Supported by the Health Research Council of New Zealand; Australian New Zealand Clinical Trials Registry number, ACTRN12620000034932.).

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Open-Source Automated Insulin Delivery in Type 1 Diabetes
Date Crossref
08/09/2022
Éditeur
Massachusetts Medical Society
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Diabetes Management and ResearchMobile Health and mHealth ApplicationsPancreatic function and diabetes

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.