Diagnostic accuracy of SARS-CoV-2 Panbio™ rapid antigen diagnostic tests in a 4,440-case clinical follow-up

6Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : hu. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Severe Acute Respiratory Syndrome Coronavirus-2 (SARS-CoV-2) Rapid Antigen Detection Testing (RADT) has been subjected to several evaluations in reference to diagnostic accuracy, ranging from small scale up to large population studies including nation-wide community-based studies. All confirmed the diagnostic accuracy of the tests which were strongly dependent upon the infection's population prevalence. In our retrospective study, parallel SARS-CoV-2 Panbio™ RADT assay, including real-time reverse transcription quantitative polymerase chain reaction (RT-qPCR) tests, were aimed to evaluate diagnostic performance regarding the rapid antigen diagnostic testing. Out of 4,440 paired tests, 609 samples tested positive using RT-qPCR, resulting in a prevalence of 13.7%. Panbio detected 251 (5.7%) positive tested samples. Overall sensitivity was 41.2% (95% CI 37.4-45.2%) and overall specificity was 99.7% (95% CI 99.4-99.8%). Positive predictive value (PPV) was 95.1% (95% CI 91.8-97.1%) and the negative predictive value (NPV) was 91.4% (95% CI 90.5-92.2%). RADT sensitivity increased with stratification in reference to the results according to PCR Cycle threshold (Ct) and presence of the symptoms considerably influenced PPV and NPV. Sensitivity in the group of Ct values ≤ 20 was 91.2%, 68.6% within the Ct range of 20-25, 47.9% in the group of Ct values between 25 and 30, and 12.6% in the group of Ct values between 30 and 35. A follow-up of the positive cases aligned with RT-qPCR testing and comparison of the general population enrolled in the testing in which the fatal cases occurred enabled us to estimate real clinical diagnostic performance regarding the SARS-CoV-2 Panbio RADT. Based upon our results, we recommend the SARS-CoV-2 Panbio RADT tests be carried out as the primary test, without parallel PCR testing, only among high population prevalence rates of the infection and to be used for symptomatic individuals with average or low severe disease developmental risk. In the case of high risk regarding the development of severe infection complications, a parallel SARS-CoV-2 RT-qPCR is needed to be carried out to attain proper diagnostic accuracy and avoid delaying appropriate medical care.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Diagnostic accuracy of SARS-CoV-2 Panbio™ rapid antigen diagnostic tests in a 4,440-case clinical follow-up
Date Crossref
02/08/2022
Éditeur
Frontiers Media SA
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

SARS-CoV-2 detection and testingBiosensors and Analytical DetectionAdvanced biosensing and bioanalysis techniques

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.