Impact of additional HEPA filter on APAP performance and CPAP pressure level in simulated sleep apnea events

1Citations signalées, ce qui n’est pas une note de qualité
4Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : ch. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background CPAP is the first line treatment of obstructive sleep apnea. Recently, the use of added filters has been debated following the field safety notice of Philips Respironics™ on potential health risks due to foam degradation used in their ventilators. However, the added resistance of filters has never been analyzed. Objectives The primary aim was to investigate the impact of four different filters on APAP mode performance with and without added unintentional air leaks (UIAL) with two simulated respiratory events. The secondary aim was to assess the pressure drop due to the increased filter resistance at different fixed CPAP pressure levels. Method This is a bench study. Performance tests were performed on a breathing simulator (ASL 5000™) with a DreamStation™ device. To assess the combined effect of UIAL, a controlled valve was added to the setup. Results Without UIAL, the algorithm was able to detect respiratory events and increase pressure level consequently. In the presence of UIAL, the device's response to simulated events was affected. In fixed CPAP mode, the median measured end-expiratory pressure was 6.2 to 10.0% (p < 0.001) below the set pressure with the additional filters. Additional UIAL severely impacted the delivered pressure with a median reduction up to 28.3% (p < 0.001) to the set pressure. Conclusion Despite a slight pressure drop, the APAP algorithm still performed with additional filters when UIAL were avoided. However, the combined effect of added filter resistance and UIAL severely impacted APAP performance and effectively delivered set pressure.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Impact of additional HEPA filter on APAP performance and CPAP pressure level in simulated sleep apnea events
Date Crossref
27/07/2022
Éditeur
Frontiers Media SA
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Obstructive Sleep Apnea ResearchInhalation and Respiratory Drug DeliveryRespiratory Support and Mechanisms

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.