Primary stability of multi-hole cups compared to plate osteosynthesis in osteoporotic anterior column and posterior hemi-transverse acetabular fractures—A biomechanical comparison

4Citations signalées — pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

INTRODUCTION: Acetabular fractures pose high demands on the surgeon and in the case of osteosynthetic treatment, anatomical reconstruction has the highest priority to achieve a good outcome. However, especially in older patients with poor bone quality, even anatomical reconstruction is no guarantee for a good clinical outcome and may nevertheless end in early osteoarthritis. Primary arthroplasty therefore has an increasing importance in the treatment of these patients. The aim of this study was to biomechanically compare fracture gap displacement and failure load as an assessment measure of the primary stability of conventional plate osteosynthesis with the treatment using a sole multi-hole cup for acetabular fractures. METHODS: Six hemi-pelvises each with anterior column and posterior hemi-transverse (ACPHT) fracture were treated with either plate osteosynthesis or a multi-hole cup. The tests were carried out in a standardised test set-up with cyclic loading in various stages between 150 N and 2500 N. The fracture gap displacement was recorded with optical 3D measuring and the failure load was determined after the cyclic measurement. RESULTS: With increasing force, the fracture gap displacement increased in both procedures. In each group there was one treatment which failed at the cyclic loading test and a bone fragment was broken out. The primary stability in arthroplasty was comparable to that of the standard osteosynthesis. CONCLUSIONS: The results found seem promising that the primary arthroplasty with a sole multi-hole cup and corresponding screw fixation achieves an initial stability comparable to osteosynthesis for typical ACPHT fractures. However, further clinical studies are needed to prove that the cups heal solidly into the bone.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Primary stability of multi-hole cups compared to plate osteosynthesis in osteoporotic anterior column and posterior hemi-transverse acetabular fractures—A biomechanical comparison
Date Crossref
27/07/2022
Éditeur
Public Library of Science (PLoS)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Pelvic and Acetabular InjuriesOrthopaedic implants and arthroplastyHip and Femur Fractures

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.