E-205 In vitroevaluation of full-length, mr-safe interventional passive catheter markers at 3T

0Citations signalées — pas une note de qualité
3Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

Introduction MRI-guided intervention provides valuable functional information such as perfusion and diffusion that standard fluoroscopic guidance lacks. Recent studies have demonstrated utility in stroke treatment for multimodality X-ray and MR imaging suites. Yet, clinical adoption of MRI-guided intervention has lagged, and an absence of appropriate tooling could be a contributing factor. Prior work described a polymeric catheter featuring radiopaque markers doped with iron oxide nanoparticles (IONPs) as low-profile, MRI-visible passive markers. Here, we further investigate susceptibility artifacts in vitro at 0° and 90° with respect to B0. Marker separation was evaluated to identify a potential format for full-length visibility at 3.0T. Methods Polymeric, 6.6 Fr guide catheter segments were built in a fabrication facility (Penumbra, Inc., Alameda, CA). Epoxy-based radiopaque ink was mixed with 9.1 wt% (w/w) iron(III) oxide (Fe2O3) nanoparticles of 20–40 nm diameter. Four circumferential marker bands ranging from 5 to 20 mm apart were applied in series starting 1 cm from the distal tip and cured at 20°C. 15-cm segments submerged in copper sulfate solution (CuSO4, 1–2 g/L) were oriented parallel (0°) and orthogonal (90°) with respect to B0. At 0°, images were acquired in coronal/sagittal slice orientations on a 3.0T MRI scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany). At 90°, images were acquired in coronal/axial orientations to match image acquisition of 0°. Phase encoding direction was maintained parallel to the catheter major axes. T2-weighted turbo spin echo (TSE) and gradient echo (GRE) sequences were collected. Maximum artifact widths were evaluated per ASTM F2119–07, using ImageJ (http://imagej.nih.gov/ij). Results Susceptibility artifacts induced by the IONP-doped markers are shown (FIG. 1). 90° orientation produced worst-case artifacts in all sequences except the axial TSE. Furthermore, some artifacts could not be discerned at 90°. 20-mm separation yielded distinct artifacts without overlapping. Conclusion Although worst-case artifacts were observed at 90°, they remained acceptable for use in guide catheters. Additionally, 20-mm marker separation could provide distinct artifacts down the length of a catheter during MRI-guided interventions, despite anatomic tortuosity. The phantom used in this study was shallow, which limited the ability to distinguish between real artifacts and those arising from the solution-air interface. Hence, accuracy of the degree of increased artifact may warrant further investigation. This study simulated worst-case orientations in vitro; more data should be collected in vivo using real-time sequences to best evaluate artifacts during clinical use. Disclosures B. Kilbride: None. C. Jordan: None. S. Ahn: 5; C; Siemens Medical Solutions USA Inc. K. Mueller: 5; C; Siemens Medical Solutions USA Inc. A. Chu: 5; C; Penumbra Inc. D. Barry: 5; C; Penumbra Inc.. T. Moore: None. A. Martin: None. M. Wilson: None. S. Hetts: None.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.

Titre Crossref
E-205 <i>In vitro</i>evaluation of full-length, mr-safe interventional passive catheter markers at 3T
Date Crossref
01/07/2022
Éditeur
BMJ Publishing Group Ltd.
Type
proceedings-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Advanced MRI Techniques and ApplicationsRadiomics and Machine Learning in Medical ImagingMRI in cancer diagnosis

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.