Quantification of Infliximab and Adalimumab in Human Plasma by a Liquid Chromatography Tandem Mass Spectrometry Kit and Comparison with Two ELISA Methods
Rattachement africain : fr. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Background: This study compared the performance of plasma infliximab and adalimumab quantification using a commercially available kit (mAbXmise kit) and mass spectrometry readout to that of two ELISA methods in patients treated for inflammatory bowel disease. Methods & results: The mAbXmise method based on liquid chromatography tandem mass spectrometry (LC-MS/MS) was linear from 2 to 100 μg/ml. It was validated according to international guidelines. Regarding cross-validation for infliximab (n = 70), the mean bias with LC-MS/MS assay was approximately threefold higher with the commercial ELISA assay compared with the in-house ELISA (-6.1 vs -1.8 μg/ml, respectively). The mean bias between the LC-MS/MS assay and in-house ELISA was -1.2 μg/ml for adalimumab (n = 35). Conclusion: The LC-MS/MS method is a powerful alternative to immunoassays to monitor concentrations of infliximab and adalimumab.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Quantification of Infliximab and Adalimumab in Human Plasma by a Liquid Chromatography Tandem Mass Spectrometry Kit and Comparison with Two ELISA Methods
- Date Crossref
- 01/06/2022
- Éditeur
- Informa UK Limited
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.